Tempsens CALsys 1500 L Скачать руководство пользователя страница 6

4. Overview

4.1 Introduction:

The ‘

CALsys 1500 L’ 

has been designed to provide stable and accurate temperature to enable

professionals   to   calibrate   Temperature   Sensing   Devices   by   comparison   method.   The

CALsys 1500 L’

model has been designed to be rugged and easily maintained. This model

provides   an   isothermal   enclosure   (ceramic   block)   in   which   the   thermocouple/RTD   can   be
calibrated   against   the   temperature   of   the   calibrator.   For   traceable   calibration   a   master
calibration sensor should be placed into the metal block alongside the unit under calibration.
The   method   is   widely   accepted   because   the   calibrator   provides   very   stable   temperature
nearing   to   its   controlled   point,   the   Master   Thermocouple   /   RTD,   which   is   calibrated   by
independent   Laboratory   traceable   to   National   standards,   compares   the   sensor   under
calibration.
The ‘

CALsys

’ models are part of wide range of Laboratory calibrators designed and made by

us. Please contact us in case you required more information about our other products.

DANGER!

 Power is present even if the front panel circuit breaker is in the OFF (0)

position.

4.2 Outline Description:

‘The calibrator controller uses a precision R type thermocouple as a controlling sensor and
controls the well temperature with 6 silicon carbide heater. To obtain and maintain a required
temperature the controller varies the power to the heater via solid-state relay. There is one
electricity driven fan which is situated under the heating chamber for cooling the heater. 

The

CALsys 1500 L dry block calibrator was designed for laboratory, moderate cost and ease of
operation.  With  proper  use   the   instrument   should  provide   continued  accurate  calibration  of
temperature sensors and devices. The user should be familiar with the safety guidelines and
operating procedures of the calibrator as described in the User’s Manual.

[!]Warning:

  Do   not   insert   impure   materials,   metal   objects  or   probes   etc.  into   the

block.

Содержание CALsys 1500 L

Страница 1: ...CALsys 1500 L CALsys 1500 L Dry block Temperature Calibrator USER MANUAL TEMPSENS INSTRUMENT I PVT LTD UNIT II...

Страница 2: ...2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 The T...

Страница 3: ...rough defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplier...

Страница 4: ...een yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside Contact us for repair Do not use the apparatu...

Страница 5: ...ncy 50 60 Hz Temperature range 500 to 1500 C Resolution 0 1 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 hours Operating Temperature 20 to 45 C Controller Specificat...

Страница 6: ...ange of Laboratory calibrators designed and made by us Please contact us in case you required more information about our other products DANGER Power is present even if the front panel circuit breaker...

Страница 7: ...ve all packing sheet from inside Insert the heaters in to the groves of SS chamber as per specified same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY Unpack the furnace carefu...

Страница 8: ...y screwdriver connect the Phase and Netural wire as per given picture Again fix the front and back plate of calsys 1500L Fix the 6 heater into the chamber and Connect all the aluminum strip in series...

Страница 9: ...ully assembled of CALsys 1500L 5 Operating Instruction 1 PID Controller 2 ON OFF switch of controller 3 Power cable RS 232 Power MCB FAN 4 RS 232 cable 5 Safety controller for over heating temperature...

Страница 10: ...switch the CALsys 1500 L ON or OFF to raise or lower the temperature The displayed temperature will gradually change until it reaches the setpoint temperature The temperature bath requires 15 to 20 m...

Страница 11: ...0L to new location NOTE USE INSULATION WOOL FOR HEAT LOSS AT FRONT SIDE OF CERAMIC BLOCK 6 Operation of Controller FRONT PANEL LAYOUT Operator Buttons 6 1 The Temperature Controller The controller has...

Страница 12: ...amp cloth and mild detergent Do not use harsh chemicals on the surface which may damage the paint Avoid knocking or dropping the calibrator If the mains supply cord becomes damaged replace it with a c...

Страница 13: ...andevaluate measuring data 1 1 Installation Install the calibration software using the installation guide file on CD ROM After installation of the software Double click the application It willopen th...

Страница 14: ...gging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVIValue Read th...

Страница 15: ...ilicon gel to protect it from humidity Warranty TEMPSENS CALsys 1500L instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are no...

Страница 16: ...TEMPSENS INSTRUMENT I PVT LTD...

Отзывы: