Tempsens CALsys 1500 L Скачать руководство пользователя страница 10

(6)

 1."R" type thermocouple for controlling and 

over temperature protection.

    

                                                                              Inner chamber of calsys 1500L

5.1 Power 

Place the calibrator on a flat surface with at least 20 inches of free space around the instrument.

Overhead clearance is required. Do not place under a cabinet or structure. Insert equilizing block

into the well with help of tong, after inserting remove the tong from equalizing block. Plug the

power cord into mains outlet of the proper voltage, frequency Typically this will be (230 VAC ±10,

50/60 Hz). Turn the “Mains” (Heater) switch located on front panel. The controller dispaly should

illuminate after 3-4 sec . After a brief self test the controller should begin normal operation. If the

unit fails to operate please check the power connection.

[!]Warning:

  Do   not   connect   or   disconnect   while   power   is   on.   Do   not   disconnect

power when furnace is hot and fans are running.
5.2 Setting the Temperature

Press “UP” or “DOWN” buttons to change the change the temperature set-point value. When the

set-point temperature is changed the controller will switch the CALsys 1500 L ON or OFF to raise or

lower the temperature. The displayed temperature will gradually change until it reaches the setpoint

temperature. The temperature bath requires 15 to 20 minutes to reach the set-point depending on

the span. Another 10 to 15 minutes is required to stabilize the temperature bath within ± 0.5ºC of

the set-point.

5.3 Operating Instruction:- 

1.

Insert the ceramic block (thermowell) into well of furnace. Block should be face to face of inner

chamber.

2.

Connect the ‘CALsys 1500L’ to a suitable power supply & Switch ON the Dry bath source and set

the desire temperature value in PID by using UP & Down key.

3.

Switch ON the MCB,back side of furnace.

4.

Insert the  reference  (Master)  temperature sensor and  UUC (Unit  under  calibration)  into the

thermowell (inconel block). 

5.

PV   (Present   value)   display   in   controller   will   gradually   rise   until   it   reaches   the   set   point

temperature. The  

controller takes some times to reach the set-point depending on the span.

Furnace is stable when PV is equal to SV (Set Value).

Note:

All other controller parameters are company set and locked. It is 

recommended not to change them.

When the source is operated at any temperature above ambient, 

the front face and plate become hot.

UP

Key

Down

Key

Содержание CALsys 1500 L

Страница 1: ...CALsys 1500 L CALsys 1500 L Dry block Temperature Calibrator USER MANUAL TEMPSENS INSTRUMENT I PVT LTD UNIT II...

Страница 2: ...2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 The T...

Страница 3: ...rough defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplier...

Страница 4: ...een yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside Contact us for repair Do not use the apparatu...

Страница 5: ...ncy 50 60 Hz Temperature range 500 to 1500 C Resolution 0 1 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 hours Operating Temperature 20 to 45 C Controller Specificat...

Страница 6: ...ange of Laboratory calibrators designed and made by us Please contact us in case you required more information about our other products DANGER Power is present even if the front panel circuit breaker...

Страница 7: ...ve all packing sheet from inside Insert the heaters in to the groves of SS chamber as per specified same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY Unpack the furnace carefu...

Страница 8: ...y screwdriver connect the Phase and Netural wire as per given picture Again fix the front and back plate of calsys 1500L Fix the 6 heater into the chamber and Connect all the aluminum strip in series...

Страница 9: ...ully assembled of CALsys 1500L 5 Operating Instruction 1 PID Controller 2 ON OFF switch of controller 3 Power cable RS 232 Power MCB FAN 4 RS 232 cable 5 Safety controller for over heating temperature...

Страница 10: ...switch the CALsys 1500 L ON or OFF to raise or lower the temperature The displayed temperature will gradually change until it reaches the setpoint temperature The temperature bath requires 15 to 20 m...

Страница 11: ...0L to new location NOTE USE INSULATION WOOL FOR HEAT LOSS AT FRONT SIDE OF CERAMIC BLOCK 6 Operation of Controller FRONT PANEL LAYOUT Operator Buttons 6 1 The Temperature Controller The controller has...

Страница 12: ...amp cloth and mild detergent Do not use harsh chemicals on the surface which may damage the paint Avoid knocking or dropping the calibrator If the mains supply cord becomes damaged replace it with a c...

Страница 13: ...andevaluate measuring data 1 1 Installation Install the calibration software using the installation guide file on CD ROM After installation of the software Double click the application It willopen th...

Страница 14: ...gging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVIValue Read th...

Страница 15: ...ilicon gel to protect it from humidity Warranty TEMPSENS CALsys 1500L instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are no...

Страница 16: ...TEMPSENS INSTRUMENT I PVT LTD...

Отзывы: