2. Before you Start

2.1 Safety Information 

Use the instrument only as specified in this manual. Otherwise, the protection provided by the 

instrument may be impaired. Refer to the safety information below and throughout the manual. 

The following definitions apply to the terms “Warning” and “Caution”.

“Warning” identifies conditions and actions that may pose hazards to the user.

“Caution” identifies conditions and actions that may damage the instrument

being used.

[!]  Warning: 

The furnace  is only to be used as described in this 

manual. It is recommended that you only use accessories provided by 
the manufacturer.

  2.2 Warning  

To avoid possible electric shock or personal injury, follow these guidelines.

This equipment must be correctly earthed.

A protective earth is used to ensure the conductive parts can not become live in the event of

a failure of the insulation.

The protective conductor of the flexible mains cable which is colored green/yellow MUST be

connected to a suitable earth.

Warning: Internal mains voltage hazard. Do not remove the panels.

There are no user serviceable parts inside. Contact us for repair.

Do not use the apparatus outside its recommended range i.e., + 500 to 1500° C.

Ensure inflammable materials, are kept away from hot parts of the apparatus, to prevent

fire risk.

Before connecting to the electricity supply, please familiarize yourself with the parts of the

calibrator with the help of operating manual.

BURN HAZARD – Do not touch the target surface of the unit.

Always replace the fuse with one of the same rating, voltage and type.

Overhead clearance is required. Do not place unit under a cabinet or other structure.

Do not use this unit for any application other than calibration work.

Completely   unattended   high   temperature   operation   is   not   recommended   for

safety reasons.

CALIBRATION EQUIPEMENT should only be used by TRAINED PERSONNEL.

Wear appropriate protective clothing.

2.3 Caution   

To avoid possible damage to the instrument, follow these guidelines.

Components   and   heater   lifetime   can   be   shortened   by   continuous   high   temperature
operation.

Содержание CALsys 1500 L

Страница 1: ...CALsys 1500 L CALsys 1500 L Dry block Temperature Calibrator USER MANUAL TEMPSENS INSTRUMENT I PVT LTD UNIT II...

Страница 2: ...2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 The T...

Страница 3: ...rough defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplier...

Страница 4: ...een yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside Contact us for repair Do not use the apparatu...

Страница 5: ...ncy 50 60 Hz Temperature range 500 to 1500 C Resolution 0 1 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 hours Operating Temperature 20 to 45 C Controller Specificat...

Страница 6: ...ange of Laboratory calibrators designed and made by us Please contact us in case you required more information about our other products DANGER Power is present even if the front panel circuit breaker...

Страница 7: ...ve all packing sheet from inside Insert the heaters in to the groves of SS chamber as per specified same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY Unpack the furnace carefu...

Страница 8: ...y screwdriver connect the Phase and Netural wire as per given picture Again fix the front and back plate of calsys 1500L Fix the 6 heater into the chamber and Connect all the aluminum strip in series...

Страница 9: ...ully assembled of CALsys 1500L 5 Operating Instruction 1 PID Controller 2 ON OFF switch of controller 3 Power cable RS 232 Power MCB FAN 4 RS 232 cable 5 Safety controller for over heating temperature...

Страница 10: ...switch the CALsys 1500 L ON or OFF to raise or lower the temperature The displayed temperature will gradually change until it reaches the setpoint temperature The temperature bath requires 15 to 20 m...

Страница 11: ...0L to new location NOTE USE INSULATION WOOL FOR HEAT LOSS AT FRONT SIDE OF CERAMIC BLOCK 6 Operation of Controller FRONT PANEL LAYOUT Operator Buttons 6 1 The Temperature Controller The controller has...

Страница 12: ...amp cloth and mild detergent Do not use harsh chemicals on the surface which may damage the paint Avoid knocking or dropping the calibrator If the mains supply cord becomes damaged replace it with a c...

Страница 13: ...andevaluate measuring data 1 1 Installation Install the calibration software using the installation guide file on CD ROM After installation of the software Double click the application It willopen th...

Страница 14: ...gging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVIValue Read th...

Страница 15: ...ilicon gel to protect it from humidity Warranty TEMPSENS CALsys 1500L instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are no...

Страница 16: ...TEMPSENS INSTRUMENT I PVT LTD...

Отзывы: