Tempsens CALsys 1500 L Скачать руководство пользователя страница 13

CAUTIONARY NOTE   

TEMPSENS PRODUCTS ARE INTENDED FOR USE BY TECHNICALLY TRAINED AND COMPETENT
PERSONNEL FAMILIAR WITH GOOD MEASUREMENT PRACTICES.
IT IS EXPECTED THAT PERSONNEL USING THIS EQUIPMENT WILL BE COMPETENT WITH THE
MANAGEMENT   OF   APPARATUS   WHICH   MAY   BE   POWERED   OR   UNDER   EXTREMES   OF
TEMPERATURE, AND ARE ABLE TO APPRECIATE THE HAZARDS WHICH MAY BE ASSOCIATED
WITH, AND THE PRECAUTIONS TO BE TAKEN WITH, SUCH EQUIPMENT.

8. Software Installation

The provided Tempsens software offers possibilities toconnect furnace temperature bath and
change set point,maximum time span, view real time graph andevaluate measuring data.

1.1 Installation

Install the calibration software using the installation guide file on CD ROM. After installation of 

the software; Double click the application. It willopen the screen of software.

1.2 Parameter in mainscreen
1.2.1 Communication

Communication  between  the  furnace  and   the  software   is  implemented  via  a  RS-232   cable
connected   between   the   furnace   and   the   PC   serial   port.   This   enables   the   acquisition   and

recording of data, as well as the transfer of commands from the software application to the
tempsens furnace. Communication can be done by clicking on connect and select correct COM

port address (fig. 1) where furnace is connected. Also user has to select type of controller
version3216 (fig. 2). Then click on CONNECT button. Shown com2 connected successfully.

Содержание CALsys 1500 L

Страница 1: ...CALsys 1500 L CALsys 1500 L Dry block Temperature Calibrator USER MANUAL TEMPSENS INSTRUMENT I PVT LTD UNIT II...

Страница 2: ...2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 The T...

Страница 3: ...rough defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplier...

Страница 4: ...een yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside Contact us for repair Do not use the apparatu...

Страница 5: ...ncy 50 60 Hz Temperature range 500 to 1500 C Resolution 0 1 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 hours Operating Temperature 20 to 45 C Controller Specificat...

Страница 6: ...ange of Laboratory calibrators designed and made by us Please contact us in case you required more information about our other products DANGER Power is present even if the front panel circuit breaker...

Страница 7: ...ve all packing sheet from inside Insert the heaters in to the groves of SS chamber as per specified same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY Unpack the furnace carefu...

Страница 8: ...y screwdriver connect the Phase and Netural wire as per given picture Again fix the front and back plate of calsys 1500L Fix the 6 heater into the chamber and Connect all the aluminum strip in series...

Страница 9: ...ully assembled of CALsys 1500L 5 Operating Instruction 1 PID Controller 2 ON OFF switch of controller 3 Power cable RS 232 Power MCB FAN 4 RS 232 cable 5 Safety controller for over heating temperature...

Страница 10: ...switch the CALsys 1500 L ON or OFF to raise or lower the temperature The displayed temperature will gradually change until it reaches the setpoint temperature The temperature bath requires 15 to 20 m...

Страница 11: ...0L to new location NOTE USE INSULATION WOOL FOR HEAT LOSS AT FRONT SIDE OF CERAMIC BLOCK 6 Operation of Controller FRONT PANEL LAYOUT Operator Buttons 6 1 The Temperature Controller The controller has...

Страница 12: ...amp cloth and mild detergent Do not use harsh chemicals on the surface which may damage the paint Avoid knocking or dropping the calibrator If the mains supply cord becomes damaged replace it with a c...

Страница 13: ...andevaluate measuring data 1 1 Installation Install the calibration software using the installation guide file on CD ROM After installation of the software Double click the application It willopen th...

Страница 14: ...gging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVIValue Read th...

Страница 15: ...ilicon gel to protect it from humidity Warranty TEMPSENS CALsys 1500L instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are no...

Страница 16: ...TEMPSENS INSTRUMENT I PVT LTD...

Отзывы: