PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions
17
18.
Use the TORX-driver with a T-10 TORX tip to secure the Eye mezzanine
board to the main board using the two screws provided in the kit. Tighten
these screws to 0.6 N·m (5.31 in-lb).
Figure 21:
Attach the Eye mezzanine board
19.
Perform steps 8 to 12 in reverse order to reinstall the Mixed Media board
back into instrument. Tighten the four Mixed Media board bracket screws
to 0.6 N·m (5.31 in-lb).
20.
Secure the back panel to the instrument. Tighten two side screws to 0.9
N·m (7.97 in-lb).
21.
Tighten the two lock nuts on the opposite side of the Mixed Media board
to 0.9 N·m (7.97 in-lb).
22.
Secure the rear panel plate. Tighten the four rear panel plate screws to 0.9
N·m (7.97 in-lb).
23.
Reinstall the rear panel hardware. Tighten the five BNC nuts to 0.9 N·m
(7.97 in-lb).
24.
Place the cover over the instrument and replace the top cover screws.
Tighten the cover screws to 0.9 N·m (7.97 in-lb).