PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions
MPI2-10 hardware upgrades
These instructions are for the qualified, trained personnel or personnel familiar
with servicing the product. Contact your nearest Telestream service center or
factory service for installation assistance.
WARNING.
Dangerous voltages may be present. To prevent electrical shock,
disconnect the power cord from the unit before starting the upgrade procedure.
Failure to do so could cause serious injury or death.
CAUTION.
Many components within the instrument are susceptible to static
discharge damage. To prevent static discharge damage, service the product
only in a static-free environment. Observe standard handling precautions for
static-sensitive devices while installing this kit. Always wear a grounded wrist
strap, grounded foot strap, and static resistant apparel while installing this kit.
Recommended tool list
Table 2:
Recommended tool list
Name
Description
9/16 inch socket nut driver
Used to removed BNC nuts.
Screwdriver handle
Accepts Torx-driver bits
T-10 TORX tip
Used for removing instrument screws.
TORX-driver bit for T-10 size screw heads.
T-15 TORX tip
Used for removing instrument screws.
TORX-driver bit for T-15 size screw heads.
Torque driver
Used to apply precisely a specific torque to
a fastener.