PRISM MPI2-10 and MPX2-10 Field Upgrade Kit Instructions
NOTE.
If power to the instrument is lost before it begins erasing the internal flash
memory, restart the firmware upgrade after the instrument reboots.
Upgrade the Firmware
1.
Copy the upgrade file with a .bin file extension from the upgrade package
onto a USB memory device. You need approximately 1 GB of available
space on the USB device.
NOTE.
If the upgrade package included a readme.txt file, read the file
before performing the upgrade. The file contains important information
about the firmware release.
2.
Insert the USB memory device into a USB port on the PRISM
monitor.
3.
On the PRISM monitor, select the
Settings
icon (
) to open the
Settings menu.
4.
Select
Utilities
to open the Utilities submenu.
5.
Select
Firmware Upgrade
. After the USB device has been
scanned, the display lists all of the files on the USB device with a
.bin file extension.
6.
Select the file in the list, and then select
Install
to start the
upgrade. (See
Error! Reference source not found.
Figure 23:
Selecting the firmware upgrade file
7.
While the upgrade is in process, the message box shown in
Figure 23
appears stating that the firmware installation is in progress.
CAUTION.
Removing the USB device or powering off the instrument before the
upgrade is complete can cause upgrade failure. To prevent upgrade failure, wait
until the
Installation Complete
message box is displayed before performing these
actions. The upgrade may take up to five minutes.
NOTE.
The upgrade may take up to five minutes.