Page 51 of 51
TVP Manual (full) March 2007 CE rev 03.doc
There has been a
momentary break in the
continuity between the
vacuum system and the
TVP
Check connection look for damaged pins
and poor connections – loose wires,
within the consumer side of the
interface connector shell.
The measured values
jump around when the
remote is connected
Voltage is being applied to
the remote connector to
system ground
Check supplied contacts are at system
ground
There is no signal back
showing defrost
complete
TVP supplied with 24VAC
for signalling but pins are
isolated zero volt relays
Check a signal is supplied which can be
switched
I only have two
available control relays
for the system
Not a problem
Since the control hierarchy places cool
over defrost. Which terminates
automatically on completion. A start
stop and a cools no cool signal are
required (this assumes that a defrost
will always follow cool) the defrost
command is made within the remote
connector
Controller displays
nonsense
Power failure
Check power supply
HV / RF electrical
interference from vacuum
plant
Check interface is earthed and
electrically isolated. Use the 10mm
earthing stud to make bond between
TVP and vacuum system so they are at
equal potential
Section 4. Tables and data
This section is intended to be a primary reference for the installation and operation of
Telemark TVP water vapor cryotraps. Details of products are subject to continuous revision
and improvement.
4.10 Description of electrical supplies, and controls electrical and system schematics
4.11 System with side panel removed showing main sub-systems
4.20 Interfacing to TVP
4.30 Principles of operation
4.41 Description of safety systems
4.43 Operational pressures and temperatures
4.50 Cooling curve of TVP series units (50 Hz)
4.60 Recommendations for the design, design and placement of cryo-coils
4.70 Recommended spares and complete spare parts list.
4.80 Materials safety data sheet
4.10 Description of electrical supplies, and controls electrical and system schematics
The full electrical circuit diagrams are provided as PDF’s on the CD-ROM – requires Adobe
Acrobat viewer 4.0 or higher.