3.3 Trouble shooting guide – problems during defrost
System defrosts slowly
Cryo-coil or refrigeration
lines have become
damaged
It is common for the refrigeration lines
to be twisted or damaged. The entire
refrigeration circuit should be free from
restrictions
Hand valves not fully open
Check defrost and suction line hand
valves have been opened
Cryo-coil has too big a
thermal mass
Redesign cryo-coil using thinner gauge
material
System momentarily
enters defrost and then
terminates to standby
Defrost requested from
standby and therefore
terminates correctly
Controller has identified
that CO or SP values are
out of range.
The CO sensor problem
System will only defrost only if coil is
cold.
Check value of CO is lower than the
defrost termination point defined in
cal2309 (default is zero
o
C.) Adjust set
point and
SP is greater than 70 psi check water
flow direction.
Check that the remote sensor
connection plug is in place if removed CI
and CO sensors will read open circuit
value grater than 200 oC
Spare sensors are fitted to system and
maybe found behind female side of
connector shell of remote temperature
sensor connection. Check with your
service center for details.
System defrosts
normally for a period
and then stops
Either the defrost or
common return hand valve
have been damaged by
operation whilst they are
cold
Consult Telemark service center, repair
requires skilled technician to correct.
3.4 Trouble shooting guide – Interface and remote control
System does not
respond to remote
interface
Has the system switched
over from local control to
remote?
Check that pin 9 is terminated against a
system ground (this is standard inside
the remote connector supplied_
Is the signal given a no volt
contact (relay) closure?
TVP system does not recognise voltage
inputs or other forms of signal
System incorrectly wired
There are several traditional interface
solutions which pre-existing vacuum
plant maybe configured to check the
TVP remote interface schematic.
Suggested wiring configurations are
shown