TVP Manual (full) March 2007 CE rev 03.doc
2.06 Installation of the Cryo-coil (Meisner trap)
The design and placement of a cryo-coil is a specialised task, which must be carried out by a
skilled vacuum practitioner. Free technical help and design service are available directly from
Telemark. Telemark also manufactures cryo-coils to order to suit your installation.
WARNING
Failing to leak test the system as a whole may result in the catastrophic release of refrigerant,
which presents a very high risk from frostbite and/or asphyxiation. See emergency shut down
procedures and the material safety data sheet for guidance.
CAUTION
Do not connect the TVP to an existing cryo-coil without ensuring that the cryo-coil will accept
the standing pressure of the TVP system (245psi) and that the whole system has been fully
leak checked. Failure to do so could damage the coil and the vacuum system.
It is assumed that the cryo-coil and associated vacuum feed-through are in their final
operating position and are leak tight to better than 2 x 10
-8
mbar/l/s. The only satisfactory
way to verify the system is leak tight to the required standard is by using a mass
spectrometer.
It is good practice for cryo-coils with attached feed-through to be installed from inside the
chamber so that there are no connections within your vacuum system, verify that moving parts
within your vacuum chamber do not interfere with the cryo-coil. Telemark supplies a range of
high quality vacuum feedthroughs, which compliment its range of refrigeration lines.
CAUTION
The cryo-coil should be more than 2 diameters away from the chamber wall or any other
object including other parts of the cryo-coil. It is bad practice to mount the cryo-coil directly
onto a metal plate. The ideal cryo-coil should have a large surface area but also be of low
mass. The cryo-coil should be capable of accepting a working pressure of 245 psig. If
chamber radiant heating is being used aluminium foil placed between the cryo-coil and the
heat source is a very effective hear shield. Radiated heat from any source in a vacuum
chamber will have a detrimental effect on your systems performance.
WARNING
Fully familiarise yourself with the main functional elements of the system before installing.
CAUTION
Refer to site guide and ensure all facilities are available before installing your TVP.