80
dB
Each dB measurement is different between the input signal and a stored relative
value, with both values converted to dBm. The dB function applies to AC voltage
and DC voltage measurements only.
Press
→
[
dB/dBm
]
→
[
On
]
and select
[
Function dB
]
to enter the interface
displayed in diagram 2- 61.
Diagram 2- 61 dB Function Interface
Table 2- 25 dB Measurement Function Menu Function Description
Function
Menu
Settings
Description
dB/dBm
On/Off
Turn on or turn off the dB or dBm function.
Function
dB
Open the dB function and the lower right corner of the
main display will show “dB”.
Ref R
Set the parameter via the direction keys: 50Ω ~ 8000Ω.
dB Ref
Value
Set the referred value of dB.
Measure
Ref Value
Set the referred value of measurement.
Done
Save all changes and return to the higher level menu.
Содержание T3DMM4-5