SeaBat T-Series Subsea Operator's Manual
Page 13
January 31, 2019
Version 7
3.4
System Shutdown Instructions
3.4.1 AUV System
Simply cutting power to the SSP can result in corrupted data files and problems with restarting
the SSP Windows® 7-based operating system. Follow the shutdown instructions in this section
carefully to prevent this from happening.
To shut down the system via the VCC:
1.
Send remote control command 1000 to the 7k Center.
Please refer to the Data Format Definition document for instructions (see
appendix
E.1 SeaBat T-Series Subsea Documentation
).
2.
Wait at least 10 seconds, and then remove power from the SSP.
To shut down the system via remote desktop:
1.
Access the SSP operating system via remote control software.
2.
Activate the 7k Center application window.
3.
Press Esc (on the keyboard). This will initiate shutdown of the sonar.
4.
Once the 7k Center application window has closed, click Start.
In the Search Programs and Files, enter: “Shutdown –s –t 1” and press Enter on the
keyboard. Wait for the remote desktop window to terminate.
5.
Once the remote desktop window has terminated, power may be removed from the
SSP.
3.4.2 ROV System
CAUTION
Ensure that the SSP is shut down correctly.
1.
Click the Application Settings button in the top right corner of the Sonar UI.
Содержание SeaBat T Series
Страница 6: ......