28
PI-MTE System Manual
Issue 6
•
User Manuals
PI-MTE System and WinView/32 Imaging Software.
•
Host Computer
Typically, the computer is user-supplied.
•
Coolant Circulator
Not required by some systems. Typically, the coolant circulator and hoses are
user-supplied.
4.3
System Requirements
4.3.1
Environmental
•
Storage temperature: <55°C
•
Operating environment: 5°C < T < 30°C
•
Relative humidity:
≤
50%; non-condensing (not applicable for open-nose)
4.3.2
Ventilation
•
ST-133
There is an internal fan located at the right side of the rear panel behind an
exhaust opening. Its purpose is simply to cool the controller electronics. This
fan runs continuously whenever the controller is powered. Air enters the unit
through ventilation openings on the side panels, flows past the warm electronic
components as it rises, and is drawn out the rear of the controller by the fan. It
is important that there be an adequate airflow for proper functioning. As long as
both the controller’s intake ventilation openings and the fan exhaust opening
aren’t obstructed, the controller will remain quite cool.
4.3.3
Coolant
WARNING!
Coolant is harmful if swallowed. Keep out of reach of
children.
A PI-MTE camera has been designed for heat removal via circulating coolant (50:50
mixture of ethylene glycol and water) for proper operation.
•
Flow Rate
2 liters/minute. Users are advised to install a flow meter to monitor the rate.
CAUTION!
!
If the flow rate is restricted or lower, then the camera can be
damaged due to overheating or it may not reach
temperature.
•
Fluid Pressure
25 psig (maximum).
Содержание Princeton Instruments PI-MTE
Страница 1: ...www princetoninstruments com PI MTE System Manual 4411 0097 Issue 6 October 1 2019...
Страница 22: ...22 PI MTE System Manual Issue 6 This page is intentionally blank...
Страница 25: ...Chapter 3 Installation 25 PI MTE System Diagram 4411 0097_0014...
Страница 26: ...26 PI MTE System Manual Issue 6 This page is intentionally blank...
Страница 58: ...58 PI MTE System Manual Issue 6 This page is intentionally blank...
Страница 66: ...66 PI MTE System Manual Issue 6 Figure 6 8 Flow Chart Safe Mode versus Fast Mode 4411 0097_0032...
Страница 90: ...90 PI MTE System Manual Issue 6 This page is intentionally blank...
Страница 94: ...94 PI MTE System Manual Issue 6 Figure B 2 PI MTE Camera 8 01 05 and later 4411 0097_0054...
Страница 95: ...Appendix B Outline Drawings 95 Figure B 3 PI MTE Camera 8 01 05 and earlier 4411 0097_0055...
Страница 102: ...102 PI MTE System Manual Issue 6 This page is intentionally blank...
Страница 112: ...112 PI MTE System Manual Issue 6 This page is intentionally blank...
Страница 117: ...This page is intentionally blank...