25
Chapter 4: Internal Shutter
The IsoPlane has an optional internal shutter. If the
shutter is installed, it is located in the Turret
Compartment and is mounted to the spectrograph
wall. Its shutter cable plugs into a keyed 2-pin shutter
connector in the floor of the IsoPlane. The external
connector is below the detector mounting plate and
can be connected via a shutter cable to a detector with
a SHUTTER connector. If a detector does not have a
SHUTTER connector (e.g., PI-MAX detectors,) the
Princeton Instruments SHC-EXT Shutter Control
Box can be used to control the opening and closing of
the internal shutter. If the control box is not available, the internal shutter can be removed.
REFERENCES:
For information about the installation and removal of the internal
shutter assembly, refer to:
•
Section 4.1.1, Install an Internal Shutter
•
Section 4.1.2, Remove an Internal Shutter
4.1 Internal Shutter Installation/Removal
An IsoPlane may be ordered with an optional internal shutter at
the entrance slit. This shutter is controlled by either:
•
A detector via a shutter cable from the detector’s Shutter
connector to the Shutter connector on the IsoPlane, or
•
A Shutter Control box (Model SHC-EXT) available from
Princeton Instruments.
This section provides procedures for the installation and removal
of an internal shutter.
4411
-014
3_000
5
4411-
0143
_0021
Содержание Everywhereyoulook IsoPlane 320A
Страница 1: ...so lane Manual ...
Страница 2: ......
Страница 14: ...10 This page is intentionally blank ...
Страница 34: ...30 This page is intentionally blank ...
Страница 46: ...42 This page is intentionally blank ...
Страница 66: ...62 This page is intentionally blank ...
Страница 74: ...70 This page is intentionally blank ...
Страница 82: ...78 ...
Страница 89: ......
Страница 90: ......