Launch the application software.
5.
Loosen the rotation adjustment screws (see
) sufficiently that you can rotate
the detector. The mounting plate holes are slotted to allow approximately 4° of
rotational positioning.
6.
If background subtraction is an available function, acquire a background to be applied
while data are acquired.
7.
Turn on the light source and begin continuous data acquisition.
8.
While viewing the data being acquired, rotate the detector until you obtain the sharpest
peaks or, if in imaging mode, the spectral lines are perpendicular in the image.
9.
Tighten the mounting screws.
10.
Confirm that the data still show sharp peaks or perpendicular spectral lines.
5.5 Focus the Spectrometer
The basic actions for focusing the spectrometer are listed below. This general procedure
assumes that you have already mounted the array detector to the exit port of the
spectrograph and that there is a light source at the entrance port. Refer to Chapter 3 of this
manual for detailed LightField and WinSpec instructions for performing this function. For
other application software, refer to that software’s manual or help file.
1.
Using a 9/64” hex wrench, unscrew the four (4) screws that secure the Micrometer
Compartment cover. See
.
Figure 5-3: Micrometer Compartment Location
2.
Set the screws aside.
44
11-0
143_0
014
M
ICROMETER
C
OMPARTMENT
Содержание Everywhereyoulook IsoPlane 320A
Страница 1: ...so lane Manual ...
Страница 2: ......
Страница 14: ...10 This page is intentionally blank ...
Страница 34: ...30 This page is intentionally blank ...
Страница 46: ...42 This page is intentionally blank ...
Страница 66: ...62 This page is intentionally blank ...
Страница 74: ...70 This page is intentionally blank ...
Страница 82: ...78 ...
Страница 89: ......
Страница 90: ......