57

Appendix B: Camera Adapters

This appendix provides detailed information about the following camera adapters available 

for use with an IsoPlane system:

Sliding Spec Flange Mount [P/N: 9000-469-01];

PyLoN Camera Mount [P/N: 9000-469];

C-Mount Camera Mount [P/N: 9000-576];

Special Spec Flange Mount [P/N: 9000-469-03].

Typically, an IsoPlane system is shipped with a 9000-469-01, Sliding Spec Flange Mount, 

although any of the others mounts listed may be selected when ordering the system.

B.1 Sliding Spec Flange Mount

The Sliding Spec Flange mount, part number 9000-469-01, is used with the following 

cameras:

PIXIS Family;

PI-MAX Family;

PyLoN-IR;

ProEM 1600.

This camera adapter is shipped with most IsoPlane systems. It enables both horizontal and 

rotational adjustments to be made to camera positioning.

Figure B-1: Typical 9000-469-01 Sliding Spec Flange Mount: Front View

4411-

0143

_0042

Содержание Everywhereyoulook IsoPlane 320A

Страница 1: ...so lane Manual ...

Страница 2: ......

Страница 3: ...0 320 Meter Focal Length Triple Grating Image Spectrograph 4411 0143 Issue 1 July 28 2021 4411 0143 IsoPlane 320A User s Manual ...

Страница 4: ...in the United States and or other countries The information in this publication is believed to be accurate as of the publication release date However Teledyne Princeton Instruments does not assume any responsibility for any consequences including any damages resulting from the use thereof The information contained herein is subject to change without notice Revision of this publication may be issue...

Страница 5: ...nal Shutter 25 4 1 Internal Shutter Installation Removal 25 4 1 1 Install an Internal Shutter 26 4 1 2 Remove an Internal Shutter 27 Chapter 5 IsoPlane Setup 31 5 1 Unpack and Inspect Equipment 31 5 2 Initial Setup Procedure 31 5 3 Array Detector Mounting Instructions 32 5 3 1 Mount an Array Detector to the IsoPlane 32 5 4 Rotational Alignment 33 5 5 Focus the Spectrometer 34 5 6 Focal Plane Adjus...

Страница 6: ... C 2 Triple Grating Turrets 63 C 3 Light Sources 63 C 4 Detectors and Accessories 65 C 5 Microscope Interface 66 C 6 Filter Wheels and Order Sorting Filter Assemblies 67 C 7 Fiber Optic Light Guides 68 C 8 Data Cables 69 Appendix D RS 232 Interface 71 D 1 Communication Configuration Settings 71 D 2 Custom Cable Information 71 Appendix E Monochromator Control Software 73 E 1 Install the Monochromat...

Страница 7: ...ternal Shutter Assembly on an IsoPlane 28 Figure 5 1 External Connector Panel 31 Figure 5 2 Typical Mounting Plate 32 Figure 5 3 Micrometer Compartment Location 34 Figure 5 4 Micrometer Locking Set Screw 35 Figure 5 5 Vertical Alignment Adjustment Set Screws 37 Figure 5 6 Horizontal Alignment Adjustment Screws 39 Figure 6 1 Typical LightField Available Devices Panel 44 Figure 6 2 Typical LightFiel...

Страница 8: ... Table C 1 Accessories Slit Assembly 63 Table C 2 Accessories Triple Grating Turret 63 Table C 3 Accessories Light Sources 63 Table C 4 Accessories Detectors 65 Table C 5 Accessories Microscope Interface 66 Table C 6 Accessories Filters 67 Table C 7 Accessories Fiber Optics 68 Table C 8 Accessories Data Cables 69 Table D 1 IsoPlane RS 232 Pin Definitions 72 ...

Страница 9: ...hs Large diameter optics and interchangeable triple grating turrets permit operation from the UV to NIR with extremely low light levels at high low or intermediate spectral resolution IsoPlane includes 32 bit microprocessor controlled scanning built in USB and RS 232 interfaces and a micrometer controlled entrance slit In addition it has a direct digital grating scan mechanism with full wavelength...

Страница 10: ... cables is compatible with the host system contact Princeton Instruments about obtaining a custom cable If you have facilities for building a custom cable refer to the RS 232 pin assignments listed in Table 1 1 Additional information in available in Appendix D RS 232 Interface on page 71 1 1 3 Grating Turret Each IsoPlane is shipped with and includes one grating turret NOTE Additional turrets are ...

Страница 11: ...nstruments products and accessories including the SpectraHub Detector Interface SpectraSense software will control and acquire data from both single channel detection systems which require scanning of a monochromator and CCD based detection systems that may or may not require repositioning of a spectrograph Entrance Slit Accessories Refer to Chapter 2 Entrance Slit Accessories on page 11 for addit...

Страница 12: ...Section C 8 Data Cables on page 69 for part number and ordering information 1 3 Connectors Ports and Major Components Figure 1 3 illustrates a typical IsoPlane spectroscope with its connectors ports and major components identified Figure 1 3 IsoPlane Connectors Ports and Major Components 4411 0143_0026 TURRET REMOVABLE MICROMETER LIGHT SOURCE INPUT ARRAY DETECTOR PORT COMPARTMENT COMPARTMENT COVER...

Страница 13: ...he IsoPlane is successfully communicating with the host computer READY This Green Yellow LED indicates the operational status of the IsoPlane spectrograph system Green Indicates that the system has been initialized Yellow Indicates that the system is busy RS 232 RS 232 connection to the host computer Refer to Appendix D RS 232 Interface on page 71 for additional information about using this interf...

Страница 14: ...10 This page is intentionally blank ...

Страница 15: ...r to accessory specific documentation for complete installation and operation information Figure 2 1 which illustrates a manual slit assembly may be helpful when mounting an accessory to the IsoPlane Figure 2 1 Typical Manual Slit Assembly Slit Position Beam Direction 0 300 7 62 mm 0 8 20 mm Micrometer Controls Slit Width from 10 Pm to 3 mm Four 4 8 32 Tapped Holes 5 16 Deep Slit Height Aperture 4...

Страница 16: ... section applies to the following manual slit models 716 1 Manual Bilateral Slit 716 MS Kinematic Slit The slit width of each bilateral slit assembly 716 type is adjustable from 0 010 mm to 3 mm 10 µm to 3 000 µm by a micrometer knob located on the top of the slit housing The micrometer knob is graduated in 0 010 mm 10 µm increments Each clockwise revolution of the micrometer knob increases the sl...

Страница 17: ...Width 200 µm 0 200 mm Figure 2 3 Typical Slit Width Micrometer Settings Width 1380 µm 1 380 mm 2 3 1 20 10 15 2 3 1 20 10 15 4411 0143_0018 MICROMETER SLEEVE SLIT WIDTH ADJUSTMENT KNOB EACH DIVISION 10 µM EACH DIVISION 250 µM EACH NUMERAL VIEW A VIEW A 1000 µM 1 MM 10 15 2 3 4411 0143_0019 ...

Страница 18: ...e bilateral slit Stepping motor controlled adjustments are made in 5 µm increments over the full range Full control of the slit including automated bandpass setting is integrated into LightField and SpectraSense software The slit motor is energized only when you send a command via the software to control it When spectrograph power is turned off the slit remains in the position to which it was last...

Страница 19: ...ratings during shipment CAUTION 1 The optical surfaces in the spectrograph are extremely delicate and can be permanently damaged by contact with solid objects as well as with most liquids and aerosols 2 DO NOT TOUCH optical surfaces with anything Wear thin protective powderless gloves to minimize damage from body oils if you accidentally touch any optical surface especially the grating 3 Avoid tal...

Страница 20: ...Using the 9 64 hex wrench remove the four Turret Compartment cover screws and remove the cover Become familiar with the grating turret and its components 3 Before removing the grating turret from the shipping container use a 3 32 hex wrench to loosen the three set screws on the bottom of the container about 1 turn 4 Remove the cover from the shipping container grasp the Center Torque Screw and tur...

Страница 21: ...ings on the grating turret assembly See Figure 3 3 Figure 3 3 Location of Hole on Turret Grating Assembly 6 Locate the vertical post on the Grating Spindle Assembly See Figure 3 4 Figure 3 4 Location of Post on Grating Spindle Assembly 4411 0143_0008 HOLE 4411 0143_0009 POST ...

Страница 22: ...urret into Turret Compartment 8 Position the grating turret on the grating spindle so the hole in the turret assembly aligns with the post in the turret assembly as shown in Figure 3 6 Figure 3 6 Turret Installed with Hole and Post Aligned 4411 0143_0010 4411 0143_0011 ALIGNED HOLE AND POST ...

Страница 23: ...d always be re started afterwards 3 2 Remove a Turret CAUTION When removing a turret be aware of and adhere to the following guidelines The optical surfaces in the spectrograph are extremely delicate and can be permanently damaged by contact with solid objects as well as with most liquids and aerosols DO NOT TOUCH optical surfaces with anything Wear thin protective powderless gloves to minimize da...

Страница 24: ...enter Torque Screw 4 Grasp the center torque screw and carefully lift the grating turret straight up off the grating spindle assembly If the turret cannot be lifted up the center torque screw may need to be loosened a bit more See Figure 3 8 Figure 3 8 Removing the Grating Turret from the Grating Compartment 4411 0143_0024 CENTER TORQUE SCREW 4411 0143_0025 ...

Страница 25: ...ify which grating turret has been installed You may however want to label the turrets for example with the spectral range used CAUTION When handling a turret be aware of and adhere to the following guidelines The optical surfaces in the spectrograph are extremely delicate and can be permanently damaged by contact with solid objects as well as with most liquids and aerosols DO NOT TOUCH optical sur...

Страница 26: ...rque screw may need to be loosened a bit more See Figure 3 10 Figure 3 10 Removing the Grating Turret from the Grating Compartment 6 Carefully place the removed grating turret assembly into the original protective container in which it was shipped Be sure to align the post and hole Tighten the center torque screw and use a 3 32 hex wrench to tighten the three 3 set screws on the bottom of the cont...

Страница 27: ...shipping container grasp the center torque screw and turn it counterclockwise until the turret can be lifted straight up 9 Locate the hole near the edge between gratings on the grating turret assembly See Figure 3 11 Figure 3 11 Location of Hole on Turret Grating Assembly 10 Locate the vertical post on the grating spindle assembly See Figure 3 12 Figure 3 12 Location of Post on Grating Spindle Ass...

Страница 28: ...ks This sets the torque on the spindle correctly NOTE This screw cannot be over tightened It is designed to slip once the correct torque has been attained 14 Place the Turret Cover on the spectrograph Replace and tighten all cover screws 15 If the IsoPlane s power supply is on restart the IsoPlane to initiate the automatic identification of the new grating turret If the power supply is off the aut...

Страница 29: ...closing of the internal shutter If the control box is not available the internal shutter can be removed REFERENCES For information about the installation and removal of the internal shutter assembly refer to Section 4 1 1 Install an Internal Shutter on page 26 Section 4 1 2 Remove an Internal Shutter on page 27 4 1 Internal Shutter Installation Removal An IsoPlane may be ordered with an optional i...

Страница 30: ...xtreme caution when handling the grating turret so as not to touch the grating surfaces 3 Mount the shutter assembly on the two socket head screws above the entrance slit If necessary back the screws out approximately five turns in order to allow the slots to slip over the mounting screws See Figure 4 1 Figure 4 1 Mounting an Internal Shutter Assembly on an IsoPlane 4 Use a 9 64 hex wrench to tigh...

Страница 31: ...control the shutter via a Princeton Instruments SHC EXT external shutter control box if one is available This is because there is no way for these detectors to keep the shutter open or to open the shutter during an acquisition Perform the following procedure to remove an internal shutter from an IsoPlane 1 Using the 9 64 hex wrench unscrew the four 4 socket head screws that secure the turret compa...

Страница 32: ...up See Figure 4 3 Figure 4 3 Disconnecting Internal Shutter 4 Use a 9 64 hex wrench to loosen but not remove the two hex head socket screws that secure the shutter housing to the wall of the IsoPlane See Figure 4 4 Figure 4 4 Mounting an Internal Shutter Assembly on an IsoPlane 4411 0143_0023 SHUTTER SHUTTER CABLE CONNECTOR 4411 0143_0022 MOUNTING SCREWS QTY 2 ...

Страница 33: ...the two mounting screws so that they are snug and will not accidentally fall out 7 Carefully install the grating turret Refer to Section 3 1 Turret Installation on page 15 for complete information CAUTION Exercise extreme caution when handling the grating turret so as not to touch the grating surfaces 8 Place the turret compartment cover on the IsoPlane 9 Install and tighten all cover screws ...

Страница 34: ...30 This page is intentionally blank ...

Страница 35: ...the IsoPlane See Figure 5 1 Figure 5 1 External Connector Panel The power supply provides 5 VDC for the control logic and 24 VDC for the motor drives 3 With the power supply switch in the off position connect the line cord provided with the IsoPlane to the power supply and the AC line The power supply will operate with line voltages between 100 and 240 VAC 50 60 Hz 4 The IsoPlane is designed to be...

Страница 36: ...s with a slotted flange or mounting holes aligned on a 3 60 bolt circle pattern There are six 6 equally spaced 10 32 tapped holes on a 3 60 bolt circle designed to accept 10 32 hex head screws See Figure 5 2 Figure 5 2 Typical Mounting Plate 5 3 1 Mount an Array Detector to the IsoPlane The array detector mounting plate is mounted to the entrance port of the IsoPlane Perform the following procedur...

Страница 37: ...flathead stainless steel screws through the shroud and into the mounting plate f Carefully mount the shroud and plate assembly onto the front of the detector and resecure the shroud with the previously removed four 4 socket head screws g Mount the plate to the IsoPlane The detector should be upright NOTE Refer to the detector s system manual for specific instructions about how to mount the array d...

Страница 38: ...Tighten the mounting screws 10 Confirm that the data still show sharp peaks or perpendicular spectral lines 5 5 Focus the Spectrometer The basic actions for focusing the spectrometer are listed below This general procedure assumes that you have already mounted the array detector to the exit port of the spectrograph and that there is a light source at the entrance port Refer to Chapter 3 of this ma...

Страница 39: ...mponents and assemblies During the manufacturing process the IsoPlane focal plane is factory aligned so that it is precisely centered on the optical axis However in some instances when a CCD camera is attached the center pixel of the CCD camera may be located slightly off center from the optical axis of the spectrograph In most applications this is not critical However if it is important to have t...

Страница 40: ... vertical alignment is off center the preferred method of correcting the alignment is ALMOST ALWAYS to adjust the vertical positioning of the image being input to the system by translating the lens fiber up or down If adjusting the input image fails to correct the vertical alignment only then should mirror adjustments be attempted 5 6 1 1 Mirror Adjustment Procedure A mirror adjustment should be p...

Страница 41: ...ise turn the Set Screw 1 8 turn clockwise A series of incremental adjustments is recommended At the end of the procedure the Cap and Set Screws should be tight The following tools are required to perform this procedure 5 32 hex wrench 3 32 hex wrench Perform the following procedure to adjust fine tune the vertical alignment 1 Remove the Micrometer Compartment cover see Figure 5 3 to access the hei...

Страница 42: ...er being used Systems equipped with a Sliding Spec Flange Mount must use the procedure described in Section 5 6 2 1 Systems Equipped with the Sliding Spec Flange Adapter REFERENCES Refer to Section B 1 Sliding Spec Flange Mount on page 57 for additional information about this adapter Systems equipped with any other camera adapter are aligned using the procedure described in Section 5 6 2 2 Systems...

Страница 43: ... Horizontal Alignment Adjustment screws MAY be reversed CAUTION When performing this procedure be aware of and adhere to the following guidelines Each adjustment made to the Cap and Set Screws should be incremental no more than approximately 1 8 turn in either direction If the Cap Screw is turned 1 8 turn clockwise turn the Set Screw 1 8 turn counterclockwise Similarly if the Cap Screw is turned 1...

Страница 44: ...Y the two screws specified in Figure 5 6 Adjustments to other screws may result in severe misalignment of the instrument If the image is still positioned too far to the right continue making incremental adjustments to these two screws until the image is properly centered 5 7 Single Point Detector Mounting Instructions Single point detectors e g a PD 438 mount directly to the optional exit slit wit...

Страница 45: ...Chapter 5 IsoPlane Setup 41 5 8 Spectrometer Calibration Refer to the application software manual or the software help for information about calibrating the data pixels with wavelengths ...

Страница 46: ...42 This page is intentionally blank ...

Страница 47: ...Control Box may be used to control the opening and closing of the IsoPlane s internal shutter Refer to Chapter 4 Internal Shutter on page 25 for additional information 6 1 Pre Operation Checks Before launching the data acquisition software i e LightField or WinSpec verify that The detector is mounted to the IsoPlane The detector and IsoPlane are connected to the computer via the appropriate commun...

Страница 48: ...s of experiment settings NOTE If the IsoPlane 320A icon is the only device in the Experiment Devices panel the Spectrograph expander will be the only expander in the list When you drag a detector into the panel the Calibration expander and all of the detector related expanders will be displayed 6 3 1 Define the Detector and Spectrograph Peform the following procedure to define a detector and spect...

Страница 49: ...eached Clicking on the icon opens the Sensor expander which is where the set temperature can be changed 4 On the Shutter expander select Shutter Mode Normal See Figure 6 2 5 Open the Spectrometer expander and select the appropriate grating See Figure 6 2 In this case the 300 g mm Blaze 750 grating has been selected and the center wavelength has been set to 500 nm for a Mercury lamp When using a bo...

Страница 50: ...N 4 With the detector mounted to the spectrograph and connected to the computer turn power on and wait for the detector to initialize 5 Launch LightField 6 Drag the icons for the desired detector and the IsoPlane into the Experiment Devices area 7 Configure the spectrograph as follows When using a mercury source configure the spectrograph to 500 nm When using a broadband source configure the spect...

Страница 51: ...all of the ROIs Alternatively acquire an image display the large data cursor and compare the vertical bar to the line shape on the screen Rotate the detector until the line shape on the screen is parallel with the vertical bar 13 After completing the rotational alignment click the Stop button 14 Tighten the four 4 mounting plate screws 15 Use a 9 64 hex wrench to remove the cover from the Micromet...

Страница 52: ... the cursor is automatically turned on to assist in identifying peaks and viewing FWHM widths Figure 6 5 shows before and after alignment graphs Figure 6 5 Typical Alignment Helper Views Rotational Alignment 18 Once properly aligned tighten the micrometer locking set screw 19 Place the Micrometer Cover on the spectrograph 20 Replace and tighten the four 4 cover screws 21 Stop data acquisition 4411...

Страница 53: ...is is the first time you have installed WinSpec 32 and a supported detector the Camera Detection Wizard will run automatically Otherwise if you installing a new detector type click the Launch Camera Detection Wizard button on the Controller CCD tab to launch the wizard 2 On the Welcome dialog leave the checkbox unselected and click Next See Figure 6 6 Figure 6 6 Typical Camera Detection Wizard Wel...

Страница 54: ...ctrograph dialog click on the Install Remove Spectrograph button Figure 6 7 Typical Define Spectrograph Dialog 3 On the Install Remove Spectrograph dialog scroll through the list of supported spectrographs select Acton 320A click Install Selected Spectrograph and click OK to return to the Define Spectrograph dialog See Figure 6 8 Figure 6 8 Typical Install Remove Spectrograph Dialogs 4411 0143_003...

Страница 55: ...dware Initialize You should hear the turret moving Figure 6 9 Typical Define Spectrograph Dialog Acton 320A Active 5 Once the spectrograph has been identified within WinSpec the next step is to align the spectrograph optics to the detector Proceed to Section 6 4 3 Alignment on page 52 4411 0143_0035 ...

Страница 56: ...s a properly aligned system Figure 6 10 Typical WinSpec Rotational Alignment View Perform the following procedure to rotationally align the system NOTE When using an LN cooled detector fill the dewar after the detector has been powered on 1 Mount a light source e g a Princeton Instruments Hg and Ne Ar Dual Switchable light source in front of the entrance slit 2 From the application s menu bar open...

Страница 57: ... 4 screws at the corners of the detector mounting plate b While watching a live display of the spectrum displayed as a graph select a peak to monitor c Carefully and slowly rotate the detector Up to 4 degrees of rotation is possible The width of the selected peak will vary from broad to narrow and back to broad The detector is propreraly aligned when it is set for the narrowest achievable width Tw...

Страница 58: ...with the cursor in all three ROIs 7 After completing the rotational alignment tighten the four mounting plate screws 8 Use a 9 64 hex wrench to remove the cover from the Micrometer Compartment on the IsoPlane 9 Use a 3 32 hex wrench to loosen the micrometer locking set screw located in the micrometer compartment See Figure 6 12 Figure 6 12 Micrometer Locking Set Screw 10 While continuously acquiri...

Страница 59: ...015 1250 P 300 300 nm I3 030 300 P 5 6 µm 300 500 nm I3 030 500 P 300 750 nm I3 030 750 P 300 1000 nm I3 030 1000 P 600 300 nm I3 060 300 P 2 8 µm 600 500 nm I3 060 500 P 600 750 nm I3 060 750 P 600 1000 nm I3 060 1000 P 600 1200 nm I3 060 1200 P 1200 300 nm I3 120 300 P 1400 nm 1200 Holographic VISa a Optimum range for this grating is from 450 nm to the upper mechanical limit I3 120 HVIS P 1200 5...

Страница 60: ...rCard and VISA accepted for purchases Electronic Orders Electronic orders accepted with proper Company Letterhead and complete contact information Fax Orders Orders may be placed by fax with written confirmation Please fax orders to FAX 978 263 5086 Please include complete name and address with these orders Written Orders Submit all written orders to our mailing address Princeton Instruments 15 Di...

Страница 61: ... is shipped with a 9000 469 01 Sliding Spec Flange Mount although any of the others mounts listed may be selected when ordering the system B 1 Sliding Spec Flange Mount The Sliding Spec Flange mount part number 9000 469 01 is used with the following cameras PIXIS Family PI MAX Family PyLoN IR ProEM 1600 This camera adapter is shipped with most IsoPlane systems It enables both horizontal and rotati...

Страница 62: ...6 445 3 223 3 000 3 206 SECTION A A 003 000 239 188 2 605 2 125 077 MODEL VIEW ISOMETRIC A 159 THRU 205 THRU 6X 7 000 3 500 4 875 3 600 BC 1 250 EQ SP ON A 10 32 UNF THRU 4 376 BC EQ SP ON A 4X 2 375 2X 070 THRU 2 56 UNC THRU ON A 2 375 BC 2 014 2 438 R 25 TYP 6 375 2 014 4X R 085 X 406 LG THRU A A A B DETAIL B SCALE 8 1 CUT V GROOVE AS SHOWN C L 60 010 4411 0143_0044 ...

Страница 63: ...n optional adapter available when ordering an IsoPlane system Figure B 4 9000 469 PyLoN Camera Adapter Mechanical Drawing and Dimensions 2 4 32 00 TYP MODEL VIEW ISOMETRIC 159 THRU 205 THRU 6X 7 000 6 375 3 500 4 875 1 250 2X R 25 7 600 2 375 3 600 BC EQ SP ON A 4 376 BC R 094 4X R 25 EQ SP ON A 4X 2 438 10 32 UNF THRU 2X 070 THRU 2 56 UNC THRU ON A 2 375 BC A A SECTION A A 2 605 003 335 239 188 0...

Страница 64: ...ane system Figure B 5 9000 576 C Mount Camera Adapter Mechanical Drawing and Dimensions 32 00 TYP 4X 2 B ON ON EQUALLY SPACED 3 600 BC FROM FAR SIDE 4X 4 201 DRILL THRU FROM FAR SIDE 375 X 110 4X R1 4 3X 21 159 DRILL THRU 2 438 10 32 UNF 2B THRU 7 000 3 500 3X 7 EQUALLY SPACED 375 X 110 7 600 4 875 3 600 BC 3 880 BC R 094 TYP EQUALLY SPACED ON 3X 7 201 DRILL THRU 2X 50 070 DRILL THRU 2 56 UNC 2B T...

Страница 65: ...rrently not a standard part Figure B 6 9000 469 03 Special Spec Flange Mount Mechanical Drawing and Dimensions 2 4 32 00 TYP 4 376 BC 7 000 6 375 3 500 4 875 1 250 2X R 25 2 375 7 600 R 094 EQ SP ON A 2 438 10 32 UNF THRU 3 600 BC 159 THRU 6X 4X 205 THRU EQ SP ON A 4X R 25 A A MODEL VIEW ISOMETRIC SECTION A A 3 000 003 335 239 188 000 010 000 2 703 2 125 4411 0143_0047 ...

Страница 66: ...62 This page is intentionally blank ...

Страница 67: ... 3 mm Supplied with interchangeable 4 0 mm and 14 0 mm high baffles Table C 2 Accessories Triple Grating Turret Part Number Description 300 483 Interchangeable Triple Grating Turret Optional second and third turrets Requires gratings Refer to Appendix A Grating Information for information about ordering gratings Table C 3 Accessories Light Sources Sheet 1 of 2 Part Number Description 110 V 60 Hz 2...

Страница 68: ...m to 2 0 µm Includes housing regulated DC power supply mounting flange variable brightness control and light collection mirror Provides output from TDS 429 TDS 429 220 Dual Light Source with 30 W Deuterium and 30 W Tungsten Halogen lamps Output wavelengths 190 nm to 2 0 µm Includes housing with light collection mirror mounting flange and power supply XS 433 XS 433 220 75 W Xenon Light Source Outpu...

Страница 69: ... 439 Universal Photomultiplier Tube Housing For standard 1 1 8 side window photomultiplier tubes Requires side window PMT and photometer power supply for operation Includes integrated light tight shutter Compatible with Acton photomultiplier tubes P1 P2 P3 Customer specified 1 1 8 side window photomultiplier tubes PD 471 Photomultiplier Tube Housing For 1 8 tubes with built in HV supply NOTE Only ...

Страница 70: ...rmoelectric cooling for operation 442 1 A Thermoelectric Cooling System For use with ID 441 InGaAs detector SpectraHub SpectraHub Data Acquisition System 20 bit A D converter single channel detector interface Requires SpectraSense software Table C 4 Accessories Detectors Sheet 2 of 2 Part Number Description Table C 5 Accessories Microscope Interface Part Number Description MicroSpec Microscope Int...

Страница 71: ...ted second order radiation Automated via SP Control Cable Can be used with the IsoPlane Includes the complete FA 2448 F Filter Set mounted in a FA 2448 filter wheel FA 2448 2 Motorized Six Position Filter Wheel Assembly Includes stepping motor and FA 2448 4 controller Enables computer indexing via an RS 232 or USB interface a Requires FA 2448 F Filter Set or Customer supplied filters which are 25 ...

Страница 72: ...elengths 400 nm to 2200 nm FC 446 010 Fixed Position Fiber Adapter A low cost solution for positioning fiberoptic bundles directly at the entrance port Accepts 10 mm diameter ferrule FC 446 020 Fiber Coupler Provides horizontal translation capability for a 10 mm diameter fiber optic bundle Accepts 10 mm diameter ferrule FC 446 021 Fiber Coupler Provides X Y Micrometer alignment adjustments Accepts...

Страница 73: ...ables are not compatible with the system being used contact Princeton Instruments for information about ordering a custom cable Table C 8 Accessories Data Cables Part Number Description CC 499 2 RS 232 Serial Cable 9 pin female to 9 pin male Length 6 feet 1 82 m 3650 USB 06 USB cable Type A to B Length 6 feet 1 82 m ...

Страница 74: ...70 This page is intentionally blank ...

Страница 75: ... configured as follows 9600 Baud 8 Data Bits No Parity 1 Start Bit 1 Stop Bit D 2 Custom Cable Information This section provides the information necessary to build a custom RS 232 cable to be used with an IsoPlane spectrograph Figure D 1 DB9 Connector Pinout Table D 1 provides the pinout information for an RS 232 cable that is to be used with an IsoPlane spectrograph NOTE The pinout information in...

Страница 76: ...RS 232 Pin Definitions DB9 Male Pin Number Signal Definition Cable Color DB9 Female Pin Number 1 No Connection 1 2 Rx Data 2 3 Tx Data 3 4 No Connection 4 5 Ground Green 5 6 No Connection 6 7 RTS 7 8 CTS 8 9 No Connection 9 ...

Страница 77: ...ludes the following subdirectories Bin The executable code for the program is stored in this subdirectory Data During program installation a Monochromator Control icon is typically placed on the desktop for launching the software If an icon is not placed on the desktop the program can be launched using one of the following methods Windows Start Programs Navitage directly to the C Program Files Pri...

Страница 78: ...lane operations which can be controled include Wavelength scanning at a pre set linear scan rate Change of scanning speeds Grating selection Rapid GOTO wavelength positioning Changing of grating turrets Jogging wavelength positioning REFERENCES Refer to the Monochromator Control software for Windows for complete information about using an IsoPlane with a computer ...

Страница 79: ...422 FOCAL PLANE 60 15 3X ADJUSTABLE FEET REF 3X 10 32 X 7 8DP IN HOUSING 2 14 54 127 5 00 305 12 00 203 8 00 17 00 432 B NOTE S B 1 WEIGHTS MEASURES APPROX 60 LBS 20 4 X 17 7 X 8 5 B 9 66 245 83 21 175 6 90 19 34 491 5 00 127 127 5 00 8 56 217 DETAIL A 4X 8 32 X 3 4 DP 32 1 25 64 2 50 3 75 95 1 69 43 4411 0143_0039 ...

Страница 80: ...A with Motorized Slit Assembly Outline Drawing 16 63 422 17 73 450 19 34 491 83 21 9 64 245 FOCAL PLANE 60 15 6 90 175 4 875 4 875 124 4 313 110 8 438 214 NOTE S 1 WEIGHTS MEASURES APPROX 60 LBS 20 4 X 17 7 X 8 5 4411 0143_0040 ...

Страница 81: ...77 Declaration of Conformity This section of the IsoPlane system manual contains the declaration of conformity for IsoPlane systems ...

Страница 82: ...78 ...

Страница 83: ...carries NO WARRANTIES EXPRESS OR IMPLIED this product against defects in materials or workmanship for a period of up to one 1 year after shipment During this period Teledyne Princeton Instruments will repair or replace at its sole option any defective parts without charge to you You must deliver the entire product to the Teledyne Princeton Instruments factory or at our option a factory authorized ...

Страница 84: ...h carry NO WARRANTIES EXPRESSED OR IMPLIED all image intensifier products for a period of one 1 year after shipment See additional Limited One 1 year Warranty terms and conditions above which apply to this warranty Responsibility for shipping charges is as described above under our Basic Limited One 1 Year Warranty X Ray Detector Limited One Year Warranty Teledyne Princeton Instruments warrants wi...

Страница 85: ...tion an authorized service center 4 Before products or parts can be returned for service you must contact the Princeton Instruments factory and receive a return authorization number RMA Products or parts returned for service without a return authorization evidenced by an RMA will be sent back freight collect 5 These warranties are effective only if purchased from the Teledyne Princeton Instruments...

Страница 86: ...of incidental or consequential damages so the above provisions may not apply to you 12 When contacting us for technical support or service assistance please refer to the Princeton Instruments factory of purchase contact your authorized Princeton Instruments representative or reseller or visit our technical support page at www princetoninstruments com Contact Information Roper Scientific s manufact...

Страница 87: ...35 G Grating availability 55 Gratings 55 I Initial setup procedure 31 Interface cables 6 L LightField application software 7 Locking set screw 35 M Manual slit adjustment 12 Mercury spectrum fluorescent lights 45 Micrometer compartment 35 locking set screw 35 MicroSpec 66 Model 718 Motorized Slit 14 Model SHC EXT 25 Monochromator Control application software 73 Motorized slit 14 Mounting a detecto...

Страница 88: ...arranties image intensifier detector 80 one year 79 one year on refurbished discontinued products 79 owner s manual and troubleshooting 81 sealed chamber 80 shutter 81 software 80 vacuum integrity 80 XP vacuum chamber 79 x ray detector 80 your responsibility 81 WinSpec application software 7 Wizard Camera Detection 49 ...

Страница 89: ......

Страница 90: ......

Отзывы: