Infrared Gas Analyzer
Operation 4
4-1
Teledyne Analytical Instruments
Operation
4.1
Introduction
Although the Model 7300A is usually programmed to your application
at the factory, it can be further configured at the operator level, or even,
cautiously, reprogrammed. Depending on the specifics of the application,
this might include all or a subset of the following procedures:
•
Setting system parameters:
•
Establish a security password, if desired, requiring Operator
to log in.
•
Establish and start an automatic calibration cycle, if desired.
•
Routine Operation:
•
Calibrate the instrument.
•
Choose autoranging or select a fixed range of analysis.
•
Set alarm setpoints, and modes of alarm operation (latching,
fail-safe, etc).
•
Program/Reprogram the analyzer:
•
Define new applications.
•
Linearize your ranges.
•
Special functions setup:
•
Set output reversal.
•
Set polarity reversal or offset output.S
•
Set gain amplification.
Before you configure your 7300A, the following default values are in
effect: RANGE/APPLICATIONS: refer to data sheet on the first page of
this manual;
Range:
Manual
Alarm Relays:
Defeated, 0.00%, HI, NOT Fail/Safe, not latching
Zero:
Auto, every 0 days 0 hours
Span:
Auto, at 10%, every 0 days, at 0 hours
Password: TAI
Содержание 7300A
Страница 4: ...iv Model 7300A Teledyne Analytical Instruments ...
Страница 12: ...1 4 1 Introduction Model 7300A Teledyne Analytical Instruments ...
Страница 26: ...2 Installation Model 7300A 2 14 Teledyne Analytical Instruments ...
Страница 36: ...3 Start up and Theory of Operation Model 7300A 3 10 Teledyne Analytical Instruments ...
Страница 80: ...5 Maintenance Model 7300A 5 10 Teledyne Analytical Instruments ...