3 Maintenance
Model 6000A
3-8 Part II
Teledyne Analytical Instruments
WARNING: DANGEROUS HIGH VOLTAGES ARE PRESENT AT
THESE TERMINALS. TRAINED PERSONNEL MUST
REMOVE THE SILKSCREEN COVER ONLY. EXER-
CISE EXTREME CAUTION.
The first strip terminal has three contacts labeled N, G and H. The
labels stand for Neutral, Ground, and Hot. This is the AC power strip
terminal. It feeds AC power to other components of the Model 6000B
System, such as the D2 lamp power supply, heater, and temperature control-
ler PCB.
The second strip terminal has four contacts labeled SHLD, SIG, GND,
MEAS and REF. This strip terminals are dedicated to the signals coming
from the photodetector amplifier. The labels stand for:
SHLD: Shield. Shield form the preamplifier cable connects to this contact.