Photometric Analyzer
Operation 3
Part I 3-17
Teledyne Analytical Instruments
####.## % SO2
4 Left=#.### SZero
The zeroing process will automatically conclude when the output is within
the acceptable range for a good zero. Then the analyzer automatically returns to
the
Analyze
mode.
3.4.1.2
Manual Mode Zeroing
Press
Zero
to enter the
Zero
function. The screen that appears allows you
to select between automatic or manual zero calibration. Use the
∆∇
keys to
toggle between
AUTO
and
MAN
zero settling. Stop when
MANUAL
appears,
blinking, on the display.
Select zero
mode: MANUAL
Press
Enter
to begin the zero calibration. After a few seconds the first of
three zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predeter-
mined rate.
####.## % SO2
Zero adj:2048 CZero
The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero
and F–Zero, use the
∆∆∆∆∆∇
keys to adjust displayed
Zero adj:
value as close as
possible to zero. Then, press
Enter
.
S–Zero starts. During S–Zero, the Microcontroller takes control as in
Auto
Mode Zeroing
, above. It calculates the differences between successive sam-
plings and displays the rate of change as
Slope=
a value in parts per million per
second (ppm/s).
####.## % SO2
Slope=#.### SZero
Generally, you have a good zero when
Slope
is less than 0.05 ppm/s for
about 30 seconds.
Once zero settling completes, the information is stored in the analyzer’s
memory, and the instrument automatically returns to the
Analyze
mode.
Содержание 6000A
Страница 4: ...iv Model 6000A Teledyne Analytical Instruments...
Страница 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Страница 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Страница 65: ...Part II 1 3 Photometric Analyzer Operational Theory 1 Teledyne Analytical Instruments 6000A Unit Only D 69023...
Страница 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...
Страница 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...