3 Operation
Model 6000A
3-16 Part I
Teledyne Analytical Instruments
Readjust the gas pressure into the analyzer until the flowrate through the
Sample Cell settles between 50 to 200 cc/min (approximately 0.1 to 0.4
SCFH).
Note: Always keep the calibration gas flow as close to the flowrate of
the sample gas as possible
3.4.1 Zero Cal
The
Zero
button on the front panel is used to enter the zero calibration
function. Zero calibration can be performed in either the automatic or manual
mode.
Make sure the zero gas is flowing to the instrument. If you get a
CELL
CANNOT BE BALANCED
message while zeroing skip to section 4.4.1.3.
3.4.1.1
Auto Mode Zeroing
Observe the precautions in sections 4.4 and 4.4.1, above. Press
Zero
to
enter the zero function mode. The screen allows you to select whether the zero
calibration is to be performed automatically or manually. Use the
∆∇
arrow
keys to toggle between
AUTO
and
MAN
zero settling. Stop when
AUTO
appears,
blinking, on the display.
Select zero
mode: AUTO
Press
Enter
to begin zeroing.
####.## %
SO2
Slope=#.### CZero
The beginning zero level is shown in the upper left corner of the display. As
the zero reading settles, the screen displays and updates information on
Slope=
in percent/second (unless the Slope starts within the acceptable zero range and
does not need to settle further). The system first does a coarse zero, shown in
the lower right corner of the screen as
CZero
, for 3 min, and then does a fine
zero, and displays
FZero
, for 3 min.
Then, and whenever Slope is less than 0.01 for at least 3 min, instead of
Slope you will see a countdown:
9 Left
,
8 Left
, and so fourth. These are software
steps in the zeroing process that the system must complete, AFTER settling,
before it can go back to
Analyze
. Software zero is indicated by
SZero
in the
lower right corner.
Содержание 6000A
Страница 4: ...iv Model 6000A Teledyne Analytical Instruments...
Страница 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Страница 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Страница 65: ...Part II 1 3 Photometric Analyzer Operational Theory 1 Teledyne Analytical Instruments 6000A Unit Only D 69023...
Страница 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...
Страница 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...