3 Operation
Model 6000A
3-24 Part I
Teledyne Analytical Instruments
If above screen displays, use the
∆∇
arrow keys to Select
AUTO
, and
press
Enter.
Press
Escape
to return to the previous Analyze Function.
3.6.3 Precautions
The Model 6000A allows a great deal of flexibility in choosing ranges for
automatic range switching. However, there are some pitfalls that are to be
avoided.
Ranges that work well together are:
•
Ranges that have the same lower limits but upper limits that differ
by approximately an order of magnitude
•
Ranges whose upper limits coincide with the lower limits of the
next higher range
•
Ranges where there is a gap between the upper limit of the range
and the lower limit of the next higher range.
Range schemes that are to be avoided include:
•
Ranges that overlap
•
Ranges whose limits are entirely within the span of an adjoining
range.
•
Ranges where the zero is suppressed, is 1-10, 1-100, etc,
however, 80-100, 90-100 is ok where the zero gas is actually
100% concentration and the calibration is inverted.
Figure 3-2 illustrates these schemes graphically.
Содержание 6000A
Страница 4: ...iv Model 6000A Teledyne Analytical Instruments...
Страница 8: ...iv Part I Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Страница 62: ...iv Part II Model 6000A Photometric Analyzer Teledyne Analytical Instruments...
Страница 65: ...Part II 1 3 Photometric Analyzer Operational Theory 1 Teledyne Analytical Instruments 6000A Unit Only D 69023...
Страница 76: ...1 14 Part II 1 Operational Theory Model 6000A Teledyne Analytical Instruments...
Страница 90: ...3 Maintenance Model 6000A 3 10 Part II Teledyne Analytical Instruments...