340CF ELSD
Section 4 Operation
4-7
Load Method Page
If the number selected has no method saved, M
ETHOD
N
OT
A
VAILABLE
will be displayed. If the non-volatile memory IC chip
is damaged, the same message will be displayed.
Note
Method Zero (0) contains a default method for normal and
reverse phase purifications on a Combi
Flash
system when the
ELSD is shipped from the factory. Default method settings are
listed in Section 4.5.
Save Method Page
Up to 8 methods can be saved. Use the
▲
or
▼
keys to select a
method number and the
MENU
|
POWER
key to save the current
conditions to the selected method number. A method includes
Spray Chamber Temperature Setpoint, Drift Tube Temperature
Setpoint, Filter Weight, Calibration Factor, and Detector Gain.
Calibration Page
20% to 100% in 1% increments
Use
▲
or
▼
keys to select an action.
Use 20% with Combi-
Flash
systems.
Push
MENU
|
POWER
key to take an action and go back Home
screen.
Push
AUTOZERO
key to keep the current state and go back Home
screen.
This feature scales the detector output. This feature is useful to
match individual detectors to each other for standardization
within a lab, or set the full scale output to a value other than 5V
or 10mV. At 100%, the signal is neither amplified nor attenuated.
LOAD METHOD:
0 1 2 3 4 5 6 7 8 9
SAVE METHOD:
0 1 2 3 4 5 6 7 8
Calibration:
100%