Section 4 Operation
4-14
Column Pre-Treatment
Chromatographic columns may introduce particles into the
mobile phase, which may lead to increased noise and blocked
fluid paths. It is recommended that the chromatographic column
be flushed with at least 10 column volumes before it is connected
to the ELSD.
4.5 Default Method
The ELSD’s default method (Method 0) is factory-set with run
parameters that typically produce optimal sensitivity across a
broad range of solvent types.
Spray Chamber (SC):
35 °C
Drift Tube (DT):
45 °C
Filter (FLT):
5
Calibration (CL):
20%
Gain (GX):
PREP
Full Scale (FS):
5V