340CF ELSD
Section 4 Operation
4-3
When the ELSD finishes the start-up process it enters the
stand-by mode. Press the
MENU
|
POWER
key to exit and enter the
HOME Screen.
Home Screen
Pushing
▲
or
▼
keys will circulate XX section to review oper-
ation condition. SC, DT, OC and ET sections display their set
points to the degree for 5 seconds when selected. Then they
display the actual temperatures to the tenth of a degree.
Pushing
MENU
|
POWER
key will bring up the Menu screen. If the
key is pushed within 5 seconds after an arrow key was pushed,
the selection at XX will be displayed. Otherwise, the timer page
will always be the first menu page displayed.
If there is an asterisk (*) at the upper right corner, pushing
MENU
|
POWER
key will bring up a screen to explain what’s wrong:
XX = SC, DT, OC, ET, RST, BLN, GAS, FLT or BFT, CL, GX,
FS
SC = Spray Chamber (setpoint: 10°C to 70°C;
readout range: -10.0°C to 120.0°C)
DT = Drift Tube (setpoint: 22°C to 120°C;
readout range: -10.0°C to120.0°C)
OC = Optical Cell (setpoint: 22°C to 70°C;
readout range: -10.0°C to 120.0°C)
ET = Exhaust Tube (setpoint: 22°C to 70°C;
readout range: -10.0°C to 120.0°C)
RST= Reset baseline to about 20mV.
BLN = Baseline reading (5000.000 – 0.000mV)
GAS = Gas (0.0 to130.0psi)
FLT or BFT = Baseline Filter
(OFF, weight 1 to 10, FLT or BFT)
CL = Calibration (default 20%, or 1V; 100% = 5V)
GX = Gain (PREP (default) or HPLC (x10 resolution)
FS = Full Scale (10mv or 5V (default))
ELSD is in Standby Mode
Diagnostics Complete
Push PWR to exit
Laser Power OK
ELS DetectionXX
20.000mV 20.0°C