Teledyne LeCroy
PCI Express Mid-Bus Probe for Summit Analyzers
Version 1.1
26
6 Dual Probe Pod Setup Using External Clocking
The steps outlined above describe a single Probe/single analyzer configuration. In a dual analyzer setup,
a second probe may need to be added depending on the type of link that is being recorded.
If the analyzer internal clock is to be used, then cable the target device to the probe pod as follows -
however, omit the 3-pin reference clock cable(s) shown in the illustration.
If external clocking is to be used, using the PE014UCA-X Ref Clock cable, connect the 3-pin reference
clock cable from the 3-pin header on the system board to the
CLK IN
port on one probe pod. Then using
the PE009UCA-X Daisy Chain Cable, connect the
CLK OUT
port on that same pod and connect it to the
CLK IN
port on the second pod. The setup is shown below.