Teledyne API - T100U Manual, Addendum to T100 Manual P/N 06807
Troubleshooting and Service
33
12. Scroll to the NORM PMT value and verify that it is approximately twice the ppb
value of the span gas.
13. Perform a zero / span calibration.
14. Check the slope and offset values and compare them to the values in Table 9-5 of
the T100 Manual.
15. Steps 7 - 14 may have to be performed more than once in order to compensate for
any over/undershooting of the PMT reading based on the adjusted HVPS drive
voltage. Best results occur after performing the PMT calibration (steps 7 - 14) at
least two consecutive times.
8.4 TECHNICAL ASSISTANCE
If this addendum and its trouble-shooting / repair sections do not solve your problems,
technical assistance may be obtained from:
Teledyne API (TAPI)
Technical Support
9970 Carroll Canyon Road
San Diego, California 92131-1106 USA
Toll-free Phone:
800-324-5190
Phone:
+1 858-657-9800
Fax:
+1 858-657-9816
Email:
Website:
Before you contact Teledyne Instruments’ Technical Support, fill out the problem report form in Appendix C,
which is also available online for electronic submission at
http://www.Teledyne-API.com/forms/p-fmapicom.asp
06840B DCN 6201