5. CALIBRATION PROCEDURES
Calibration of the T100U should be performed according to the procedures described in
the T100 Manual. However, delivering span and zero gases for the lower ranges that the
T100U is designed for can be difficult. For best results when calibrating the T100U, wait
one hour
for the instrument to stabilize when delivering zero and span gases before
pressing the zero and span buttons.
Attention must be paid to the quality of the gasses,
the level of contaminants in the gases as well as the history and conditioning of the gas
delivery components. Only Teflon or glass should be used for any “wetted” surfaces
that the calibration gasses contact. All delivery system components should be
conditioned by running span gas for a minimum of four hours before conducting actual
span calibrations.