Math FFT
Eliminating Aliases
To eliminate aliases, try the following remedies:
Turn the Horizontal
Scale
knob to set the sample rate to a faster setting. Since
you increase the Nyquist frequency as you increase the sample rate, the
aliased frequency components appear at their proper frequency. If too many
frequency components are shown on the screen, you can use the FFT Zoom
option to magnify the FFT spectrum.
If you do not need to view frequency components above 20 MHz, set the
Bandwidth Limit option to On.
Put an external
fi
lter on the source signal to bandwidth limit the source
waveform to frequencies below that of the Nyquist frequency.
Recognize and ignore the aliased frequencies.
Use zoom controls and the cursors to magnify and measure the FFT spectrum.
Magnifying and Positioning an FFT Spectrum
You can magnify and use cursors to take measurements on the FFT spectrum. The
oscilloscope includes an FFT Zoom option to magnify horizontally. To magnify
vertically, you can use the vertical controls.
Horizontal Zoom and
Position
The FFT Zoom option lets you horizontally magnify the FFT spectrum without
changing the sample rate. Zoom factors are X1 (default), X2, X5, and X10. At
zoom factor X1, and with the waveform centered in the graticule, the left graticule
line is at 0 Hz and the right graticule line is at the Nyquist frequency.
When you change the zoom factor, the FFT spectrum is magni
fi
ed about the
center graticule line. In other words, the axis of horizontal magni
fi
cation is the
center graticule line.
Turn the Horizontal
Position
knob clockwise to move the FFT spectrum to the
right. Push the
Set to Zero
button to position the center of the spectrum at the
center of the graticule.
Vertical Zoom and Position
The channel vertical knobs become vertical zoom and position controls for their
respective channels when displaying the FFT spectrum. The Vertical Scale knob
provides zoom factors of X0.5, X1 (default), X2, X5, and X10. The FFT spectrum
is vertically magni
fi
ed about the M marker (math waveform reference point on the
left edge of the screen).
Turn the Vertical
Position
knob clockwise to move the spectrum up for the source
channel.
68
TPS2000B Series Digital Oscilloscope User Manual
Содержание TPS2012B
Страница 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Страница 2: ......
Страница 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Страница 8: ......
Страница 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Страница 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Страница 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Страница 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Страница 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Страница 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...