Reference
The maximum sample rate is 1 GS/s for oscilloscope models with a bandwidth
of 100 MHz and 2 GS/s for the 200 MHz model. At 100 ns and faster settings,
this sample rate does not acquire 2500 points. In this case, a Digital Signal
Processor interpolates points between the sampled points to make a 2500 point
waveform record.
Peak Detect.
Use Peak Detect acquisition mode to detect glitches as narrow as
10 ns and to limit the possibility of aliasing. This mode is effective when the
Horizontal Scale setting is 5
μ
s/div or slower.
Peak Detect acquisition intervals (1250)
• Sample points displayed
Peak Detect mode displays the highest and lowest acquired voltage in each interval.
NOTE.
If you set the Horizontal Scale setting to 2.5
μ
s/div or faster, the acquisition
mode changes to Sample because the sample rate is suf
fi
ciently fast to make Peak
Detect unnecessary. The oscilloscope does not display a message to tell you
that the mode was changed to Sample.
When there is enough waveform noise, a typical peak detect display shows large
black areas. The oscilloscope displays this area with diagonal lines to improve
display performance.
Typical peak detect display
TPS2000B peak detect display
Average.
Use Average acquisition mode to reduce random or uncorrelated noise
in the signal you want to display. Data is acquired in sample mode, then a number
of waveforms are averaged together.
Select the number of acquisitions (4, 16, 64, or 128) to average for the waveform.
Run/Stop Button.
Push the
Run/Stop
button when you want the oscilloscope to
continuously acquire waveforms. Push the button again to stop the acquisition.
94
TPS2000B Series Digital Oscilloscope User Manual
Содержание TPS2012B
Страница 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Страница 2: ......
Страница 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Страница 8: ......
Страница 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Страница 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Страница 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Страница 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Страница 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Страница 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...
Страница 184: ...Index Z Zoom 58 FFT 68 Horiz menu 104 Window Zone 104 105 158 TPS2000B Series Digital Oscilloscope User Manual...