Managing TPSBAT Battery Packs
Maintaining Battery Packs
Observe and note the run time that a new fully-charged battery provides for
powering your product. You can use this new battery run time as a basis to
compare run times for older batteries. The run time of your battery will vary
depending on the product’s con
fi
guration and the applications that you run.
Routinely check the charge status of the battery.
Carefully monitor batteries that are approaching the end of their estimated life.
Consider replacing the battery with a new one if the battery run time drops
below about 80% of the original run time, or the battery charge time increases
signi
fi
cantly.
Follow the storage requirements if you store or do not use a battery for an
extended period. If you do not follow the storage requirements, and the
battery has no charge remaining when you check it, you can consider the
battery to be damaged. Do not attempt to recharge it or to use it. Replace
it with a new battery.
General Charging Guidelines
Continuous Charging
You do not need to continuously charge (trickle charge) Li-Ion battery packs to
maintain full operating capacity between jobs. However, a Lithium-ion battery
pack will self-discharge during non-use. To achieve the longest operating time,
charge your battery pack before use. If you plan to store battery packs, read the
speci
fi
c instructions to do this. (See page 91,
Charging Temperature
Charge the battery packs within an ambient temperature range of 0 °C to +40 °C
(+32 °F to +104 °F). Charging battery packs outside of this range can damage
cells or cause them to leak. Charging ef
fi
ciency is best between 0 °C and +30 °C
(+32 °F and +86 °F) with less than 80% relative humidity.
Discharging Temperature
The battery packs are rated to discharge between -10 °C and +50 °C (-14 °F and
+122 °F) with less than 80% relative humidity. Discharging outside of this range
can cause damage. Battery discharge capacity drops signi
fi
cantly at temperatures
below 0 °C (+32 °F) and above +40 °C (+104 °F).
Cold temperatures adversely affect the normal electrochemical reactions within
a battery pack, reducing the available capacity. Although Li-Ion battery packs
can discharge down to -10 °C (-14 °F) without damage, the available capacity is
signi
fi
cantly reduced below 0 °C (+32 °F). You can minimize this loss by keeping
the battery packs above 0 °C (+32 °F) prior to and during use.
86
TPS2000B Series Digital Oscilloscope User Manual
Содержание TPS2012B
Страница 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Страница 2: ......
Страница 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Страница 8: ......
Страница 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Страница 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Страница 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Страница 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Страница 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Страница 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...
Страница 184: ...Index Z Zoom 58 FFT 68 Horiz menu 104 Window Zone 104 105 158 TPS2000B Series Digital Oscilloscope User Manual...