Performance Verification
4- 6
TPS2000 Series Digital Storage Oscilloscopes Service Manual
10.
Disconnect the test setup.
11.
Repeat steps 1 through 10 until all input channels have been checked.
This test checks the time base accuracy.
1.
Set up the oscilloscope using the following steps:
Press menu button
Select menu option
Select setting
DEFAULT SETUP
—
—
CH 1 (or whatever channel is
currently being tested).
Voltage Probe
1X
2.
Connect the oscilloscope to the time mark generator as shown below.
Digitizing oscilloscope
50
Ω
feedthrough
terminator
BNC cable
Time mark
generator
Output
3.
Set the time mark generator period to
10 ms
.
4.
Set the oscilloscope VOLTS/DIV to
500 mV/div
.
5.
Set the oscilloscope Main SEC/DIV to
1 ms/div
.
6.
Press
SET LEVEL TO 50%
.
7.
Use the vertical POSITION control to center the test signal on screen.
8.
Use the horizontal POSITION control to set the position to
10.00 ms
.
9.
Set the oscilloscope SEC/DIV to
250 ns/div
.
Check Sample Rate and
Delay Time Accuracy
Содержание TPS2000 Series
Страница 6: ......
Страница 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...