Maintenance
TPS2000 Series Digital Storage Oscilloscopes Service Manual
6--25
When the oscilloscope is on and operating properly, the PROBE COMP output
should generate a square wave, approximately 5 V in amplitude, at a 1 kHz
frequency. Use the oscilloscope and set the Attenuation switch to 10X on the
P2100 probe to probe this output.
As shown in the table below, there are two cases of defects. Either the 1 kHz
signal is on and the IO and Acquisition boards are both active and functionally
capable of performing an acquisition, or one of those two modules is not
performing correctly. While a large number of different possible frequencies
exist, all involve detected failures on the Acquisition or IO boards due to failure.
Signal at PROBE COMP
Possible problem
1 kHz signal, no display
Troubleshoot the display
Non-1 kHz signal
Troubleshoot the Acquisition board
No signal
Troubleshoot the power supply
Failures that occur during the power-up diagnostics or normal oscilloscope
operation place an entry in the error log.
Accessing the error log.
The error log list can be accessed anytime by pushing the
UTILITY
button and then selecting
Options
>
Error Log
. Table 6--4 lists the
error codes and probable causes.
Table 6--4: List of error codes
Number
ID
Description
Probable cause
0
Battery
Battery 1 alarm
Battery hardware failure;
replace battery If errors
1
Battery
Battery 2 alarm
replace battery. If errors
persists, contact Tektronix
2
Battery
PIC Communication
May occur sporadically
during normal operation. If
errors occur repeatedly a
3
Battery
PIC HW Failure
errors occur repeatedly, a
hardware failure may exist.
Contact Tektronix.
0
Diagnostics
Diagnostics internal error
Contact Tektronix
4
Diagnostics
CH1 diagnostics failed
Acquisition board or power
supply failure
5
Diagnostics
CH2 diagnostics failed
Acquisition board or power
supply failure
7
Diagnostics
ADG421 diagnostics failed
Acquisition board or power
supply failure
PROBE COMP Output
Using the Error Log
Содержание TPS2000 Series
Страница 6: ......
Страница 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...