Maintenance
TPS2000 Series Digital Storage Oscilloscopes Service Manual
6--21
See
boot up
screen?
Errors?
Front panel
problems?
Run probe check
on each channel
Put scope or
frequency counter
on probe comp
Replace
Acquisition
board
1 kHz?
Verify LCD
connector and
cable values
Good
voltages?
Replace
Acquisition board
Replace display
or display cable
Yes
No
Probe
check ok?
Replace
Acquisition
board
Connect
external trigger
to probe comp
Set trigger
source to EXT
Press set 50%
Trigger
indicator ON?
No
Yes
Replace
Acquisition
board
Good scope
Yes
No
Yes
Yes
No
Yes
No
Yes
Turn on power
Does screen
lighten?
Yes
From Figure 6--11
To Figure 6--11
No
A
B
To Figure 6--10
C
To Figure 6--10
D
No
No
Figure 6--9: Oscilloscope troubleshooting tree (1 of 4)
Содержание TPS2000 Series
Страница 6: ......
Страница 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...