Performance Verification
TPS2000 Series Digital Storage Oscilloscopes Service Manual
4- 7
10.
Check that the rising edge of the marker crosses the center horizontal
graticule line within
±
2 divisions of center graticule.
NOTE
. One division of displacement from graticule center corresponds to
a 25 ppm time base error.
11.
Disconnect the test setup.
This test checks the edge trigger sensitivity for all input channels.
1.
Set up the oscilloscope using the following steps:
Press menu button
Select menu option
Select setting
DEFAULT SETUP
—
—
CH 1 (or whatever channel is
currently being tested).
Voltage Probe
1X
TRIGGER
Mode
Normal
ACQUIRE
Sample
—
MEASURE
Source
An unchecked channel
Type
Pk-Pk
2.
As shown below, connect the oscilloscope channel selected in the table to the
leveled sine wave generator.
Digitizing oscilloscope
Leveled
sine wave
generator
50
Ω
feedthrough
terminator
BNC cable
Output
3.
Set the oscilloscope VOLTS/DIV to
500 mV/div
.
4.
Set the oscilloscope SEC/DIV to
25 ns/div
.
Check Edge Trigger
Sensitivity
Содержание TPS2000 Series
Страница 6: ......
Страница 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Страница 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...