Functional check
3.
Set the probes offset to 1 V and the vertical scale to 200 mV/div.
4.
Set the function generator to 1 Vpp amplitude 1 V offset and the function to a 10 MHz sine wave.
5.
The instrument should show a 1 Vpp sine wave on the screen centered around 1 V.
6.
Open the vertical channel menu and change the coupling mode to DC reject.
7.
Set the vertical offset to 0 V.
8.
Con
fi
rm that the signal is centered around 0 V.
TPR1000 and TPR4000 User Manual