Probing principles
Probing principles
Follow these helpful hints to make probing easier and noise free.
Ground lead length
When you are probing a circuit, you should
always use as short a ground lead as
possible between the probe head and circuit
ground. (See the illustration for the effects of
lead length on waveform distortion.)
The series inductance added by the probe
tip and ground lead can result in a resonant
circuit; this circuit may cause parasitic ringing
within the bandwidth of your oscilloscope.
TPR1000 and TPR4000 User Manual