Installation
Menu button
1.
Press the probe
Menu
button to display
the Probe Control screen on the
oscilloscope.
2.
Use the touch-screen buttons on the
instrument to set the probe parameters.
3.
Press the probe
Menu
button again to
close the Probe Control screen.
AutoZero
We recommend that you run the probe
AutoZero routine:
After the 20 minute warm-up period
When the operating temperature of the
probe changes by ±5 °C
1.
Press the probe
Menu
button to display
the Probe Control screen on the
oscilloscope.
2.
Short the probe tip to ground.
3.
Press the
AutoZero
button on the
instrument to execute the AutoZero
routine.