Performance veri
fi
cation
3.
Set the oscilloscope settings as follows:
Control
Setting
Vertical
500 mV/div
Horizontal
10
μ
s/div
Record Length
1000
Acquire menu
Sample
Trigger position
50%
Trigger slope
Rising Edge
Trigger level
+0.5 V
Input impedance
1 M
Ω
Measure
Amplitude
4.
Measure the low and high levels and record the results in the test record.
1 V AC coupled into 75
Ω
.
5.
On the SPG8000, set the Word Clock output to 1 V (AC):
a.
Press the
AES
button.
b.
Press the up (
▲
) arrow button until you see
AES WORD CLOCK
OUTPUT
.
c.
Press the right (
►
) arrow button until you see
1 Volt (AC)
, and then press
the
ENTER
button.
6.
Set the oscilloscope settings as follows:
Control
Setting
Vertical
500 mV/div
Horizontal
200
μ
s/div
Record Length
1000
Acquire menu
Sample
Trigger position
50%
Trigger slope
Rising Edge
Trigger level
+0.0 V
Input impedance
1 M
Ω
Measure
Amplitude
7.
Measure the level and record the result in the test record.
58
SPG8000 Speci
fi
cations and Performance Veri
fi
cation