Performance veri
fi
cation
9.
Look at the three ramps on the waveform monitor display. The larger of the
two ramps should each have 16 equal steps.
NOTE.
View the waveform display in full screen mode for easiest viewing.
10.
Use the Gain function on the waveform monitor to expand the signal to 10X.
11.
Now check that the shallowest ramp has 16 equal steps.
12.
Record Pass or Fail in the test record.
13.
Check that the RAMP HPF calibration setting has not changed. If it has,
then restore the original value.
14.
Press the
BACK
button to exit the Calibration menu.
15.
Repeat this procedure for the Black 2 and Black 3 outputs of the SPG8000.
Black amplitude and offset
test (SD)
Perform this procedure to check that Black signal output amplitude and offset for
SD signals are adjusted to within speci
fi
cation. This test should be performed in
factory mode. (See page 19,
1.
Connect the BNC-to-Banana-plug adapter to the voltmeter.
2.
Connect the BNC T to the adapter.
3.
Connect a 75
Ω
precision terminator to one end of the BNC T connector.
4.
Connect a 75
Ω
BNC cable to the other end of the BNC T connector.
5.
Connect the other end of the cable to the
BLACK 1
output on the SPG8000.
Figure 8: Setup for Black amplitude and offset test (SD)
SPG8000 Speci
fi
cations and Performance Veri
fi
cation
45