Performance veri
fi
cation
Set up the digital signal analyzer.
CAUTION.
Electrostatic discharge can damage the oscilloscope modules. To
prevent damage, always work in a static free environment and discharge the
static voltage from your body by wearing a grounded antistatic wrist strap while
handling these modules.
1.
Install the 80A03 output cable into the Channel 1/2 slot of the oscilloscope.
2.
Install the 80E04 sampling head into the 80A03 adapter and connect the two
using the SMA cables.
3.
Install the TCA-75 into the left port of the 80A03.
4.
Install the TCA-BNC into the right port of the 80A03.
Figure 27: Set up the digital signal analyzer
5.
If needed, press the Default Settings button on the oscilloscope.
6.
Press the Channel 1 button on the 80E04 sampling head to activate Channel 1.
98
SPG8000 Speci
fi
cations and Performance Veri
fi
cation