Removal and Installation Procedures
TDS6000 Series Service Manual
6- 13
1.
Remove the left and right covers:
See Figures 6--4 and 6--5, pages 6--14
and 6--15.
H
Trim (all)
H
Bottom cover
2.
Orient the oscilloscope:
Set the oscilloscope so its rear is on the work
surface and the front of the oscilloscope facing you.
NOTE
. All mounting screw holes are indicated by a star etched around the
mounting hole.
a.
Remove the eleven T-15 Torx screws that secure the covers to the top
and both sides of the chassis.
b.
Remove the seven T-15 Torx screws that secure the covers to the bottom
of the chassis.
c.
Pull the bottom-right cover down and slide to the right to remove from
the oscilloscope. Pull the top-left cover upward and slide to the left to
remove from the oscilloscope.
CAUTION.
Take care not to bind or snag the covers on the oscilloscope internal
cabling as you remove or install.
3.
Reinstallation:
Do in reverse steps a through c to reinstall the cabinet covers.
Covers
Содержание TDS6404
Страница 4: ......
Страница 14: ...Table of Contents x TDS6000 Series Service Manual...
Страница 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Страница 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Страница 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Страница 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Страница 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Страница 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Страница 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Страница 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Страница 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Страница 232: ...Options 7 4 TDS6000 Series Service Manual...
Страница 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Страница 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...