Performance Tests
4- 88
TDS6000 Series Service Manual
Table 4- 10: Clock recovery input frequencies and baud rates (Cont.)
Input frequency
Recovered clock Baud rate
477.5 MHz
1950 Mbaud
462.5 MHz
1850 Mbaud
462.5 MHz
1757 Mbaud
439 MHz
1850 Mbaud
312.5 MHz
1250 Mbaud
310 MHz
1240 Mbaud
155 MHz
620 Mbaud
155 MHz
579 Mbaud
147 MHz
620 Mbaud
109 MHz,
462 Mbaud
115 MHz
439 Mbaud
77.5 MHz
310 Mbaud
39 MHz
156 Mbaud
19.5 MHz
78 Mbaud
9.75 MHz
39 Mbaud
4.875 MHz
19.5 Mbaud
2.438 MHz
9.75 Mbaud
1.219 MHz
4.876 Mbaud
609.5 kHz
2.438 Mbaud
304.8 kHz
1.219 Mbaud
Содержание TDS6404
Страница 4: ......
Страница 14: ...Table of Contents x TDS6000 Series Service Manual...
Страница 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Страница 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Страница 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Страница 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Страница 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Страница 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Страница 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Страница 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Страница 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Страница 232: ...Options 7 4 TDS6000 Series Service Manual...
Страница 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Страница 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...