Incoming Inspection
TDS5000 Series User Manual
1- 27
Equipment
required
One passive 10X oscilloscope probe (P5050 or equivalent)
Prerequisites
None
1.
Initialize the oscilloscope:
Push the front-panel
DEFAULT SETUP
button.
2.
Hook up the signal source:
Connect the oscilloscope probe from the probe
compensation terminal to the CH 1 input as shown in Figure 1--8.
TDS5000 oscilloscope
Connect probe from PROBE
COMP terminal to CH 1 input
Figure 1- 8: Setup for trigger test
3.
Set up the oscilloscope:
Push the front-panel
AUTOSET
button.
4.
Verify that the main trigger system operates:
Confirm that the following
statements are true.
H
The trigger level readout for the A (main) trigger system changes with
the front-panel TRIGGER
LEVEL
knob.
H
The front-panel TRIGGER
LEVEL
knob can trigger and untrigger the
square-wave signal as you rotate it. (Leave the signal untriggered).
H
Pushing the front-panel TRIGGER
LEVEL
knob sets the trigger level to
the 50% amplitude point of the signal and triggers the signal that you
just left untriggered. (Leave the signal triggered.)
Trigger Test
Содержание TDS5054
Страница 5: ......
Страница 21: ...Preface xvi TDS5000 Series User Manual...
Страница 39: ...Installation 1 18 TDS5000 Series User Manual...
Страница 165: ...Triggering 3 98 TDS5000 Series User Manual...
Страница 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Страница 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Страница 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Страница 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Страница 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Страница 339: ...Glossary Glossary 12 TDS5000 Series User Manual...
Страница 361: ...Index Index 22 TDS5000 Series User Manual...