Appendix E: Optimizing Measurement Accuracy
E- 4
TDS5000 Series User Manual
Overview
Control elements and resources
To calibrate probes
Optimize gain
and offset
accuracy
2.
Connect the probe to the oscilloscope and the probe tip
to the probe compensation terminals (see
Connecting
Probes for Calibration
on page E--3).
3.
If calibrating a passive probe, first perform the
Compensating Low Frequency Probes
procedure on
page E--5.
4.
Click the VERT toolbar button to open the Vertical Setup
control window.
5.
Click the Probe Cal button to open the Probe Calibration
control window.
6.
Select the oscilloscope channel to which the probe is
attached.
7.
Click Clear Probecal to erase any previous calibration
data.
8.
Click the Calibrate Probe button.
9.
Check the Probe Status readout.
Initialized
indicates
that the probe has not been calibrated on the selected
channel.
Pass
indicates that the probe has been
calibrated on the selected channel.
Note:
Probes without a TEKPROBE interface cannot be
calibrated.
For further
assistance
10.
Access the oscilloscope online help.
See page 3--209 to learn about using online help.
Содержание TDS5054
Страница 5: ......
Страница 21: ...Preface xvi TDS5000 Series User Manual...
Страница 39: ...Installation 1 18 TDS5000 Series User Manual...
Страница 165: ...Triggering 3 98 TDS5000 Series User Manual...
Страница 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Страница 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Страница 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Страница 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Страница 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Страница 339: ...Glossary Glossary 12 TDS5000 Series User Manual...
Страница 361: ...Index Index 22 TDS5000 Series User Manual...