TDS5000 Series User Manual
B- 1
Appendix B: Automatic Measurements Supported
This appendix provides a list of all supported measurements and their defini-
tions. An illustration showing the levels used to take measurements is also
included.
Table B- 1: Supported measurements and their definitions
Name
Definition
Amplitude
Voltage measurement. The high value less the low value measured over the entire waveform or
gated region.
gated region.
Amplitude = High--Low
Area
Area measurement (Voltage over time measurement). The area over the entire waveform or
gated region in volt-seconds. Area measured above ground is positive; area below ground is
negative.
Cycle Area
Voltage over time measurement. The area over the first cycle in the waveform, or the first cycle
in the gated region, in volt-seconds. Area measured above ground is positive; area below
ground is negative.
Burst Width
Timing measurement. The duration of a burst. Measured over the entire waveform or gated
region.
Cycle Mean
Voltage measurement. The arithmetic mean over the first cycle in the waveform or the first cycle
in the gated region.
Cycle RMS
Voltage measurement. The true Root Mean Square voltage over the first cycle in the waveform
or the first cycle in the gated region.
Delay
Timing measurement. The time between the MidRef crossings of two different traces or the
gated region of the traces.
Fall Time
Timing measurement. Time taken for the falling edge of the first pulse in the waveform or gated
region to fall from a High Ref value (default = 90%) to a Low Ref value (default =10%) of its
final value.
Frequency
Timing measurement for the first cycle in the waveform or gated region. The reciprocal of the
period. Measured in hertz (Hz) where 1 Hz = 1 cycle per second.
High
The value used as 100% whenever High Ref, Mid Ref, and Low Ref values are needed (as in fall
time and rise time measurements). Calculated using either the min/max or the histogram method.
The
min/max
method uses the maximum value found. The
histogram
method uses the most
common value found above the mid point. Measured over the entire waveform or gated region.
Low
The value used as 0% whenever High Ref, Mid Ref, and Low Ref values are needed (as in fall
time and rise time measurements). May be calculated using either the min/max or the histogram
method. With the min/max method it is the minimum value found. With the histogram method, it
refers to the most common value found below the midpoint. Measured over the entire waveform
or gated region.
Maximum
Voltage measurement. The maximum amplitude. Typically the most positive peak voltage.
Measured over the entire waveform or gated region.
Mean
Voltage measurement. The arithmetic mean over the entire waveform or gated region.
Содержание TDS5054
Страница 5: ......
Страница 21: ...Preface xvi TDS5000 Series User Manual...
Страница 39: ...Installation 1 18 TDS5000 Series User Manual...
Страница 165: ...Triggering 3 98 TDS5000 Series User Manual...
Страница 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Страница 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Страница 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Страница 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Страница 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Страница 339: ...Glossary Glossary 12 TDS5000 Series User Manual...
Страница 361: ...Index Index 22 TDS5000 Series User Manual...