Triggering
3- 76
TDS5000 Series User Manual
Overview
Control elements and resources
To trigger on transition time
Set the
transition time
6.
The threshold levels and the delta time setting
determine the transition time (slew rate) setting. The
level settings determine the voltage component of slew
rate (Volts/Second). Click the Upper Level or Lower
Level button and use the multipurpose knobs or pop-up
keypad to set the values for the upper and lower levels.
You can set the level(s) to a value appropriate to either
the TTL or ECL logic families by clicking TTL or ECL on
the Level pop-up keypad.
Hint:
To use the trigger level marker to set the threshold
levels, open the Display Setup control window, and click
Long on the Objects tab.
7.
To finish specifying the slew rate (transition time), set
the time component by clicking Time and using the
multipurpose knob or keypad to set the delta time value.
8.
Click Trigger When Transition Time < to trigger when the
transition time is less than the time you set above. Click
Trigger When Transition Time > to trigger when the
transition time is greater than the time you set above.
If you select Trigger When Transition Time > (greater
than) and the oscilloscope does not trigger, the pulse
edge may be too fast. To check the pulse edge speed,
switch to edge triggering. Trigger on the pulse edge and
determine the time the edge takes to travel between the
upper and lower levels that you set for this trigger. The
oscilloscope cannot transition trigger on pulse edges
that traverse between threshold levels in 600 ps or less.
Also, to reliably trigger on transition time, a pulse must
have a width of 7.5 ns or more. A pulse of less width
may trigger on the wrong slope or not trigger at all.
Switch to edge triggering and check the pulse width if
you can’t trigger on transition time as expected.
Set mode and
holdoff
9.
Mode and holdoff can be set for all standard trigger
types. Refer to
Select the trigger mode
on page 3--56
and
Set holdoff
on page 3--59. To learn more about
trigger mode and holdoff, see
Trigger Modes
on
page 3--51 and
Trigger Holdoff
on page 3--52.
For mode and holdoff setup, see
Select the trigger
mode
on page 3--56 and
Set holdoff
on
page 3--59.
Содержание TDS5054
Страница 5: ......
Страница 21: ...Preface xvi TDS5000 Series User Manual...
Страница 39: ...Installation 1 18 TDS5000 Series User Manual...
Страница 165: ...Triggering 3 98 TDS5000 Series User Manual...
Страница 181: ...Displaying Waveforms 3 114 TDS5000 Series User Manual...
Страница 201: ...Measuring Waveforms 3 134 TDS5000 Series User Manual...
Страница 305: ...Appendix B Automatic Measurements Supported B 4 TDS5000 Series User Manual...
Страница 319: ...Appendix D Menu Bar Commands D 12 TDS5000 Series User Manual...
Страница 327: ...Appendix E Optimizing Measurement Accuracy E 8 TDS5000 Series User Manual...
Страница 339: ...Glossary Glossary 12 TDS5000 Series User Manual...
Страница 361: ...Index Index 22 TDS5000 Series User Manual...