Performance Tests
TDS 510A Performance Verification and Specifications
1–57
1. Install the test hookup and preset the instrument controls:
a. Hook up test signal:
Connect one of the 50
W
cables to CH 1. See Figure 1–21.
Connect the other end of the cable just installed to the female
BNC-to-clips adapter. See Figure 1–21.
Connect the red clip on the adapter just installed to the PROBE
COMPENSATION SIGNAL on the front panel; connect the black
clip to PROBE COMPENSATION GND. See Figure 1–21.
Digitizing Oscilloscope
50
Coaxial Cable
Female BNC to
Clip Adapter
Black Lead
to GND
Figure 1–21: Initial Test Hookup
b. Initialize the oscilloscope:
Press save/recall SETUP.
Press the main-menu button Recall Factory Setup.
Press the side-menu button OK Confirm Factory Init.
c.
Modify the initialized front-panel control settings:
Set the horizontal SCALE to 200
m
s.
Press SET LEVEL TO 50%.
Use the vertical POSITION knob to center the display on screen.
Press SHIFT; then press ACQUIRE MENU.
Press the main-menu button Mode; then press the side-menu button
Average.
Содержание TDS 510A
Страница 3: ......
Страница 4: ......
Страница 12: ...Preface viii TDS 510A Performance Verification and Specifications...
Страница 13: ...Performance Verification...
Страница 14: ......
Страница 18: ...Performance Verification Procedures 1 4 TDS 510A Performance Verification and Specifications...
Страница 32: ...Performance Tests 1 18 TDS 510A Performance Verification and Specifications...
Страница 96: ...Performance Tests 1 82 TDS 510A Performance Verification and Specifications...
Страница 97: ...Specifications...
Страница 98: ......
Страница 106: ...Nominal Traits 2 8 TDS 510A Performance Verification and Specifications...
Страница 112: ...Specifications 2 14 TDS 510A Performance Verification and Specifications...