Performance Tests
1–26
TDS 510A Performance Verification and Specifications
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
Set the output of a DC calibration generator to 0 volts.
Connect the output of a DC calibration generator through a
dual-banana connector followed by a 50
W
precision coaxial cable to
one side of a BNC T connector. See Figure 1–5.
Connect the Sense output of the generator through a second
dual-banana connector followed by a 50
W
precision coaxial cable to
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–5.
DC Calibrator
50
Coaxial Cables
Dual Banana to
BNC Adapters
BNC T
Connector
Digitizing Oscilloscope
Output
Sense
Figure 1–5: Initial Test Hookup
b. Initialize the oscilloscope:
Press save/recall SETUP.
Press the main-menu button Recall Factory Setup.
Press the side-menu button OK Confirm Factory Init.
c.
Modify the default settings:
Press SHIFT; then press ACQUIRE MENU.
Press the main-menu button Mode; then press the side-menu button
Average 16.
2. Confirm input channels are within limits for DC accuracy at maximum offset
and position: Do the following substeps — test CH 1 first, skipping substep
2a since CH 1 is already selected from step 1.
Содержание TDS 510A
Страница 3: ......
Страница 4: ......
Страница 12: ...Preface viii TDS 510A Performance Verification and Specifications...
Страница 13: ...Performance Verification...
Страница 14: ......
Страница 18: ...Performance Verification Procedures 1 4 TDS 510A Performance Verification and Specifications...
Страница 32: ...Performance Tests 1 18 TDS 510A Performance Verification and Specifications...
Страница 96: ...Performance Tests 1 82 TDS 510A Performance Verification and Specifications...
Страница 97: ...Specifications...
Страница 98: ......
Страница 106: ...Nominal Traits 2 8 TDS 510A Performance Verification and Specifications...
Страница 112: ...Specifications 2 14 TDS 510A Performance Verification and Specifications...