Performance Tests
1–44
TDS 510A Performance Verification and Specifications
CHECK that the Lower Limit readout, after the oscilloscope stops
triggering, is within 1
m
s to 3
m
s, inclusive.
Enter time on test record.
Use the keypad to return the Lower Limit to 500 ns and reestablish
triggering.
Press the side-menu button Upper Limit; then use the general
purpose knob to slowly decrease the Upper Limit readout until
triggering stops.
CHECK that the Upper Limit readout, after the oscilloscope loses
triggering, is within 1
m
s to 3
m
s, inclusive.
Enter the time on test record.
4. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.
Equipment
Required
One DC calibration generator (Item 9)
One BNC T connector (Item 7)
Two 50
W,
precision coaxial cables (Item 5)
Prerequisites
See page 1–15.
DC Calibrator
50
W
Coaxial Cables
Dual Banana to
BNC Adapters
BNC T
Connector
Digitizing Oscilloscope
Figure 1–15: Initial Test Hookup
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
Set the output of the DC calibration generator to 0 volts.
Check Accuracy,
Trigger-Level or
Threshold, DC Coupled
Содержание TDS 510A
Страница 3: ......
Страница 4: ......
Страница 12: ...Preface viii TDS 510A Performance Verification and Specifications...
Страница 13: ...Performance Verification...
Страница 14: ......
Страница 18: ...Performance Verification Procedures 1 4 TDS 510A Performance Verification and Specifications...
Страница 32: ...Performance Tests 1 18 TDS 510A Performance Verification and Specifications...
Страница 96: ...Performance Tests 1 82 TDS 510A Performance Verification and Specifications...
Страница 97: ...Specifications...
Страница 98: ......
Страница 106: ...Nominal Traits 2 8 TDS 510A Performance Verification and Specifications...
Страница 112: ...Specifications 2 14 TDS 510A Performance Verification and Specifications...