TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Instruction Manual
5- 3
TCPA300 and TCPA400 Performance Verification
This section covers the performance verification for the TCPA300 and TCPA400
Current Probe Amplifiers. Refer to the list of equipment required in Table 5--3.
Power on the amplifier and test equipment and let them warm up for 20 minutes.
Before starting these procedures, photocopy the appropriate test record beginning
on page 5--16 to record the performance test results. The recommended
calibration interval is one year.
Equipment Required
The performance verification procedures require the test equipment listed in
Table 5--3. The test equipment must meet or exceed the specifications listed. You
may need to modify the test procedures if you do not use the recommended
equipment.
Table 5- 3: Required test equipment
Qty
Item
Description
Recommended
equipment
1
Oscilloscope
500 MHz bandwidth
Tektronix TDS5000
1
Leveled Sine Wave Generator
3 MHz to 100 MHz
Wavetek 9100 with >250 MHz Oscilloscope
Calibration Option
1
Digital Multimeter
DCV: 0.1% Accuracy, 5
1
/
2
digit resolution
Keithley 2700
1
Current Source
DCA: 0.1% Accuracy, 0 to
±
100 mA,
floating outputs
Fluke 5100 or HP 6612C
1
Current Source
DCA: 0.1% accuracy, 0 to
±
15 A
Wavetek 9100
1
Square Wave Generator
ACV 1.5 Vp-p, 28 Hz square wave
Wavetek 9100
1
HF Current Loop
50
Ω
, BNC Connector
Tektronix part number 015-0601-50
1
Precision Termination
50
Ω,
±
0.1%, 0.5 W
Tektronix part number 011-0129-00
2
BNC Cables
50
Ω
, 0.76 m (30 in) long
Tektronix part number 012-0117-00
1
Cable
TEKPROBE Interface Cable
Tektronix part number 012-1605-00
2
BNC to Dual Banana Adapters
Tektronix part number 103-0090-00
1
DC Current Loop
5 turns 18 AWG coated wire
Refer to page 5--4
1
DC Current Loop
50 turns 13 AWG coated wire
Refer to page 5--4
1
Calibration Adapter
TCPA Calibration Adapter
Tektronix part number 174-4765-00
1
Current Probe
1
TCPA300/400-compatible probe
Tektronix TCP303, 305, or 312 (TCPA300) or
TCP404XL (TCPA400)
1
Required for functional test only.
Содержание TCPA300 SERIES
Страница 4: ......
Страница 8: ...Table of Contents iv TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 12: ...Table of Contents viii TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 18: ...Service Safety Summary xiv TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 20: ...Preface xvi TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 21: ...Getting Started...
Страница 22: ......
Страница 37: ...Operating Basics...
Страница 38: ......
Страница 45: ...Reference...
Страница 46: ......
Страница 64: ...Application Notes 3 18 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 71: ...Specifications...
Страница 72: ......
Страница 86: ...Specifications 4 14 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 88: ......
Страница 89: ...Performance Verification...
Страница 90: ......
Страница 135: ...Adjustment Procedures...
Страница 136: ......
Страница 138: ...Adjustment Procedures Overview 6 2 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 146: ...TCP305 and TCP312 Adjustments 6 10 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 157: ...Maintenance...
Страница 158: ......
Страница 160: ...Customer Maintenance 7 2 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 181: ...Replaceable Parts...
Страница 182: ......
Страница 190: ...Amplifier Replaceable Parts 8 8 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 198: ...Current Probes Replaceable Parts 8 16 TCPA300 400 Amplifiers and TCP300 400 Series Current Probes Instruction Manual...
Страница 199: ...Glossary and Index...
Страница 200: ......